Unit
IN 280 XT
IN 280 XTW
IN 590 XTW
Pump level
Recommended cooling water
temperature
°C
---
15
15
---
Recommended cooling water
pressure difference
bar
---
3
3
---
Minimum differential pressure of
cooling water
Á
bar
---
0.8
0.8
---
Maximum cooling water pressure
difference
bar
---
5
10
---
Maximum cooling water pressure bar
---
10
10
---
À
Pump level 4
Á
only approved for a max. cooling water temperature of 20 °C
Table 56: Cooling output
Unit
IN 1590 XTW
IN 1590 XTW
Pump level
Cooling output at (ambient tem-
perature 20 °C)
at 50 Hz
at 60 Hz
200 °C
kW
18.50
18.50
8
100 °C
kW
18.50
18.50
8
20 °C
kW
18.50
18.50
8
10 °C
kW
15.00
16.00
8
0 °C
kW
11.50
12.20
8
-10 °C
kW
8.70
9.00
8
-20 °C
kW
8.50
9.00
4
-30 °C
kW
8.50
9.00
4
-40 °C
kW
7.50
8.50
4
-50 °C
kW
6.00
7.00
4
-60 °C
kW
4.00
5.20
4
-70 °C
kW
2.20
3.40
4
-80 °C
kW
0.90
1.50
4
-90 °C
kW
0.35
0.50
2
Cooling medium
---
Water
Water
---
Cooling water flow rate
l/min
26.0
26.0
---
Cooling water temperature
minimum/maximum
°C
10/30
10/30
---
Recommended cooling water tem-
perature
°C
15
15
---
V6
Integral Process Thermostats and High-Temperature Thermostats
173 / 198
Содержание IN 130 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198...
Страница 182: ...15 9 Declaration of Conformity and certificates V6 Integral Process Thermostats and High Temperature Thermostats 182 198...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198...
Страница 199: ......