Protective equipment:
n
Safety glasses
n
Protective gloves
n
Protective work clothing
The drain taps and drain nozzles are located on the right hand side of all
devices.
1.
Allow the device and heat transfer liquid to cool or warm up to room
temperature.
2.
Screw the hose nozzle with a union nut (EOA 078) onto the drain
nozzle (3/8" exterior thread). There are two drain nozzles from the
Integral 550 XT and higher.
3.
Insert a hose on the hose nozzle for draining. Place the hose in a
suitable container to collect the heat transfer liquid.
It may be necessary to drain the device several times if the
filling volume is high.
4.
Open the drain tap. To do this, turn the tap counterclockwise.
5.
Allow the device to run empty.
6.
After the device has run empty, remove the hoses from the external
consumers. Drain the heat transfer liquid in the hoses into a suitable
container.
7.
After draining close the drain tap or taps.
The high-temperature thermostat is in standby mode.
1.
Screw the hose nozzle with a union nut (EOA 078) onto the drain
nozzle (3/8" exterior thread).
2.
Insert a hose on the hose nozzle for draining. Place the hose in a
suitable container to collect the heat transfer liquid.
3.
Press the Enter key to open the menu.
4.
Select the menu items
Fill mode
Drain
on
off.
The high-temperature valve in the device is set in such a way that
the bypass line also runs empty.
5.
Allow the device to run empty.
6.
Select the menu items
Fill mode
Drain
off
off.
The high-temperature valve closes.
Fig. 89
Fig. 90
Draining the high-temperature thermostat
V6
Integral Process Thermostats and High-Temperature Thermostats
154 / 198
Содержание IN 130 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198...
Страница 182: ...15 9 Declaration of Conformity and certificates V6 Integral Process Thermostats and High Temperature Thermostats 182 198...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198...
Страница 199: ......