Table 45: Cleaning fluids
Suitable cleaning fluid
for heat transfer liquid
Ethanol (solvent)
It is essential to observe the relevant safety
measures for handling ethanol!
Kryo 20
Kryo 51
Kryo 65
Kryo 70
Kryo 95
Ultra 350
Water
Kryo 30
11.3
Draining the cooling water circuit
This section is relevant for:
n
Water-cooled devices
To avoid damage during storage, the cooling water circuit must be drained
completely.
Protective equipment:
n
Safety glasses
n
Protective gloves
n
Protective work clothing
The device is in standby mode.
1.
Shut down cooling water supply.
2.
Unscrew the hose for the cooling water inlet from the device.
3.
Remove the filter strainer from the water inlet. If necessary, use
pointed pliers or large tweezers.
4.
Clean the filter strainer.
5.
Insert cleaned filter strainer.
6.
Press the Enter key to open the menu.
7.
Select the menu items
Fill mode
Water valve
open
on the
operating unit.
The water valve opens.
8.
Drain the cooling water circuit completely. Blow compressed air
through the cooling water circuit.
9.
Select the menu items
Fill mode
Water valve
auto
(automatic)
on the operating unit. This is the default setting.
V6
Integral Process Thermostats and High-Temperature Thermostats
156 / 198
Содержание IN 130 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198...
Страница 182: ...15 9 Declaration of Conformity and certificates V6 Integral Process Thermostats and High Temperature Thermostats 182 198...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198...
Страница 199: ......