LTI TruSense S300 Series Integration Manual 3nd Edition
Page 39
Copyright © [2016] Laser Technology, Inc. All Rights Reserved. Unauthorized duplication, in whole or in part, is strictly prohibited.
Command:
Example:
8A5!
5<CR><LF>
The sensor
’
s SDI-12 address changed from
„
8
‟
to
„
5
‟
.
“aM!” Start Measurement defined by Target Mode “aXHn!” or Serial Command $DM
This command is used to request measurement(s). The
“
aD0!
”
command will then be used to retrieve
the data when it
’
s ready.
Command:
Example:
8M!
80025<CR><LF>
Sensor 8 will have 5 measurements in 2 seconds.
followed by: (in approximately 2 seconds)
8<CR><LF>
Service Request
–
the measurements are ready.
“aM5!” Start Measurement for “Temperature”
This command is used to request measurement(s) for the
“
Temperatur
e”
data. The
“
aD0!
”
command will
then be used to retrieve the data when it
’
s ready.
Command:
command
!
= SDI-12 command termination character
ttt
= the specified time, in seconds, until the sensor will have data
ready
n
= the number of measurements being accumulated
Example:
8M5!
80015<CR><LF>
The sensor will have 5
“
Temperatur
e”
data in 1
second. followed by: (in approximately 1 second)
8<CR><LF>
Service Request
–
the measurements are ready.
“aMCn!” Start Specified Measurement with CRC
These commands are used to request specified measurement(s)
defined by “aXHn!” or Serial Command $DM
and temperature data and include the CRC when it
’
s retrieved. The
“
aD0!
”
command will then be used to
retrieve the data with the included CRC when it
’
s ready. These commands are the same as the
“
aM!
”
or
“aM0! and
“
aM5!
” c
ommands except that when the measurement data is retrieved, it will include a CRC.
Command:
aAb!
Response:
b<CR><LF>
where:
a
= specified sensor
’
s current SDI-12 address
A
= Change Address SDI-12 command
b
= sensor
’
s new SDI-12 address (
„
0
‟
to
„
9
‟
,
„
A
‟
to
„
Z
‟
,
„
a
‟
to
„
z
‟
)
!
= SDI-12 command termination character
aM!
Response:
atttn<CR><LF>
where:
a
= specified sensor's current SDI-12 address
M
= Start Measurement for
“
First #1
”
measurement data SDI-12
command
!
= SDI-12 command termination character
ttt
= the specified time, in seconds, until the sensor will have data
ready
n
= the number of measurements being accumulated
aM5!
Response:
atttn<CR><LF>
where:
a
= specified sensor
’
s current SDI-12 address
M5
= Start Measurement for
“
Temperatur
e”
measurement data
SDI-12
aMCn!
Response:
atttnn<CR><LF>
where:
a
= specified sensor
’
s current SDI-12 address
MCn
= Start Measurement for specified measurement data with CRC
SDI-12