LTI TruSense S300 Series Integration Manual 3nd Edition
Page 12
Copyright © [2016] Laser Technology, Inc. All Rights Reserved. Unauthorized duplication, in whole or in part, is strictly prohibited.
Work Sheet for setting TruSense S330 for installation in 4-20mA Loop
Sensor Serial Number: _____________ Sensor Location: ______________________________________________
Measurement unit
$MU
Question
1
Meters, two decimals?
$MU,m,2
Meters, three decimals?
$MU,m,3
Feet, two decimals?
$MU,f,2
Feet, three decimals?
$MU,f,3
ASCII Command
$MU,______,_____
ASCII Command to send
$MU,<Line1>
Current Loop $FT
Question
2
Measurement Span Height?
3
Instrument Offset Height?
______________
4
Distance Span Calculation
<Line5> + <Line 6> =___________
5
4mA Current boundary?
Distance Span or Instrument offset
Value: ______________
6
20mA Current boundary?
Distance Span or Instrument offset
Value: ______________
7
Error Current?
0 = Loop out of bounds error current 24mA
1 = Loop out of bounds error current 3.5mA
ASCII Command
$FT,_____,_____,0,_____,1
ASCII Command to send
$FT,<Line 5 >,<Line 6>,0,<Line7>,1
Liquid Setting $OS
Question
8
What Filter Type?
Smooth Surface = 1
Turbulent Surface = 2
9
Measurement per second?
Value 1 to 14: ______
10
Running Average?
Value 0=off 2-30 = on:_______
ASCII Command
$OS,_____,_____,0,_____
ASCII Command to send
$OS,<Line8 >,<Line 9>,0,<Line 10>
Error Reporting $NE
Question
11
Error Reporting?
On = 0
Off = 1
ASCII Command
$NE,_______
ASCII Command to send
$NE,<Line 11>
Consecutive Errors $CE Question
12
Consecutive Errors?
No Turbulance = 3
Slight Turbulance = 10
Moderate Turbulance = 100
Very Turbulant = 255
ASCII Command
$CE,_______
ASCII Command to send
$CE,<Line 12>
Target Mode $DM
Question
13
Target Mode?
Open Vessel = 5
Stilling Well = 7
ASCII Command
$DM,________
ASCII Command to send
$DM,<Line13>