background image

CHAPTER 3 

OPERATION 

3-3 

 

DC-24 and DC-25 operation manual 

 

3.3.

 

RS-485 parameters. 

3.3.1.

 

Parameter 1 for RS-485: Address of the display. 

 

It allows to configurate the address of the display in the RS-485 network. Value between 

0 and 99. 

3.3.2.

 

Parameter 2 for RS-485: Serial line configuration. 

The parameters of the serial line are codified in the following table: 

 

Code 

Baud Rate 

Data bits 

Parity 

Stop bits 

01 

4800 Bauds 

7 bits 

No parity 

02 

9600 Bauds 

7 bits 

No parity 

03 

19200 Bauds 

7 bits 

No parity 

04 

4800 Bauds 

8 bits 

No parity 

05 

9600 Bauds 

8 bits 

No parity 

06 

19200 Bauds 

8 bits 

No parity 

07 

4800 Bauds 

7 bits 

Even 

08 

9600 Bauds 

7 bits 

Even 

09 

19200 Bauds 

7 bits 

Even 

10 

4800 Bauds 

8 bits 

Even 

11 

9600 Bauds 

8 bits 

Even 

12 

19200 Bauds 

8 bits 

Even 

13 

4800 Bauds 

7 bits 

Odd 

14 

9600 Bauds 

7 bits 

Odd 

15 

19200 Bauds 

7 bits 

Odd 

16 

4800 Bauds 

8 bits 

Odd 

17 

9600 Bauds 

8 bits 

Odd 

18 

19200 Bauds 

8 bits 

Odd 

19 

4800 Bauds 

7 bits 

No parity 

20 

9600 Bauds 

7 bits 

No parity 

21 

19200 Bauds 

7 bits 

No parity 

22 

4800 Bauds 

8 bits 

No parity 

23 

9600 Bauds 

8 bits 

No parity 

24 

19200 Bauds 

8 bits 

No parity 

25 

4800 Bauds 

7 bits 

Even 

26 

9600 Bauds 

7 bits 

Even 

27 

19200 Bauds 

7 bits 

Even 

28 

4800 Bauds 

8 bits 

Even 

29 

9600 Bauds 

8 bits 

Even 

30 

19200 Bauds 

8 bits 

Even 

31 

4800 Bauds 

7 bits 

Odd 

32 

9600 Bauds 

7 bits 

Odd 

33 

19200 Bauds 

7 bits 

Odd 

34 

4800 Bauds 

8 bits 

Odd 

35 

9600 Bauds 

8 bits 

Odd 

36 

19200 Bauds 

8 bits 

Odd 

Table 1: RS-485 codes for configuration 

 

 

Содержание DC-24S

Страница 1: ...1600K23B TETRALEC E I S L USER MANUAL DC 24 and DC 25 Thermometer and hygrometer...

Страница 2: ......

Страница 3: ...set 3 1 3 1 1 Displays with several sensors 3 1 3 2 Programming parameters 3 1 3 2 1 Modify parameters RS 485 Ethernet or Wifi 3 2 3 3 RS 485 parameters 3 3 3 3 1 Parameter 1 for RS 485 Address of the...

Страница 4: ...display 4 8 4 4 3 Send values to the display Displays without sensor 4 9 4 5 ModBus TCP PROTOCOL 4 10 4 5 1 Modbus TCP protocol 4 10 4 5 2 Read data from the display 4 10 4 5 3 Send temperature and h...

Страница 5: ...ensors and displays its state once the display is initialized The temperature and humidity sensor is placed in a case that permits a very easy installation using the support included in the package Th...

Страница 6: ...een depending on the model Viewing distance max 50 meters Text Silkscreen Environmental Conditions Operation Temperature 20 to 60 C Storage temperature 30 C to 70 C Humidity 5 95 RH non condensing Max...

Страница 7: ...he weight of the DC 24S and DC 25S is 5kg The weight of the DC 24D and DC 25D is 7kg 1 4 Dimensions of the DC 24 and DC 25 1 4 1 Dimensions of the DC 24S and DC 25S C 520 320 40 22 1 4 2 Dimensions of...

Страница 8: ...r Insert the screws leaving their head outside for 4mm Hang the display using the upper side zig zag 2 Screwed Remove the methacrylate cover Place the 4 supplied plugs at the distance showed in the dr...

Страница 9: ...of the DC 24D and DC 25D The DC 24D and DC 25D displays must be hanged using the rings the display has Cables or chains can be used The power cables and the sensor can be fixed to the mounting elemen...

Страница 10: ......

Страница 11: ...rom dust and jets of water Displays DC 24 and DC 25 should not be installed in places with an illumination level in excess of 1000 lux Neither should the display be placed in direct sunlight as visibi...

Страница 12: ...on part protected by a cap that permits the humidity to pass through but not water The connexion cable enters through a gland The set must not be manipulated to keep the IP65 sealing Slot for mounting...

Страница 13: ...s due to doors 2 Placing it in climate control systems outputs 3 Proximity to fridges 4 Walls with direct sunlight In case the sensor cable must be extended the colours of the cables must be respected...

Страница 14: ......

Страница 15: ...Displays with several sensors In the displays with several sensors the displayed value is the mean value of the sensed by the connected sensors The display detects the connected sensors and calculates...

Страница 16: ...with the digital point indicates the parameter number and the two right digits indicate its value The flashing digit is the one the user can modify with the key To enter the menu main the key pressed...

Страница 17: ...09 19200 Bauds 7 bits Even 1 10 4800 Bauds 8 bits Even 1 11 9600 Bauds 8 bits Even 1 12 19200 Bauds 8 bits Even 1 13 4800 Bauds 7 bits Odd 1 14 9600 Bauds 7 bits Odd 1 15 19200 Bauds 7 bits Odd 1 16 4...

Страница 18: ...wing protocols 0 ASCII Really simple protocol that uses ASCII characters 1 ModBus RTU ASCII This uses the ModBus protocol but the temperature and humidity data are coded in ASCII in the same block 2 M...

Страница 19: ...is parameter performs two functions Left digit Sensor control 0 Disable sensor information 1 Enable sensor information Right digit Communication protocol The user can select between the following prot...

Страница 20: ......

Страница 21: ...There can be 99 devices connected to the net with the corresponding amplifiers every 30 devices Using this protocol forces the display to work on slave mode This means that the display waits for a me...

Страница 22: ...wing Header Address Code Temp Space Hum End of block Characters High Low P T T T T H H Carry return Decimal 64d 80d 84d 32d 13d Hexadecimal 40h 50h 54h 20d 0Dh The address returned is the one of the d...

Страница 23: ...84d 49d 56d 44d 54d 32d 52d 55d 13d Hexadecimal 40h 31h 35h 56h 54h 31h 38h 2Ch 36h 20h 34h 37h 0Dh Response from the display The response message from the display is the following Header Address Code...

Страница 24: ...and 01h The bytes 5 and 6 correspond to the number of words of the data field In the data field the characters PT must be placed so the display know the user is requesting data In the previous exampl...

Страница 25: ...umber of words of the data field In the data field the characters VT must be placed followed by the temperature and humidity data At the end a 00h character must be sent so the number of byte sent is...

Страница 26: ...Reg L Data Low CRC High CRC 06h 00h 00h 00h C5h 48h 7Bh To write the humidity value the data must be sent to register 0001h 40002 The data format must be 16 bit word format In the following example t...

Страница 27: ...cture is detailed for a humidity request Addr Func Start Addr H Start Addr L Num Reg H Num Reg L Low CRC High CRC 03h 03h 00h 03h 00h 01h 75h E8h If the block is correctly received the display respond...

Страница 28: ...4 4 2 Read values from the display To read the current temperature and humidity the following frame must be sent Code End of Block Characters P T Carry return Decimal 80d 84d 13d Hexadecimal 50h 54h 0...

Страница 29: ...imal 56h 54h 20h 0Dh Response block in protocols TCP IP and UDP IP Example The response block from the display is the following In the ModBus TCP protocol the field Data is empty Code End of Block Cha...

Страница 30: ...0h 00h 00h 06h 01h 03h 00h 02h 00h 01h Response from the display The response frame from the display is the following In this example the identifier is 0862h and the read temperature is 20 5 C CDh Ide...

Страница 31: ...20 4 C and 42 Identifier Protocol Number bytes Unit Function 01h 0Ah 00h 00h 00h 0Dh 01h 10h Data 00h 00h 00h 01h 32h 30h 2Eh 34h 20h 34h 32h 4 5 3 Send temperature and humidity In the repeater displ...

Страница 32: ...rs Identifier Protocol Num bytes Unit Function Start address Num Registers 00h 37h 00h 00h 00h 11h 01h 10h 00h 00h 00h 05h Function 10h mode Word Sending temperature and humidity can be sent using the...

Страница 33: ...on the XPort 03 to which you wish to assign the address Then press Assign IP and follow the instructions IMPORTANT All units are dispatched from the factory with the same IP address Therefore to confi...

Страница 34: ...and click OK on the bottom of the page 4 2 1 UDP IP configuration Select Channel 1 Connection in the left menu Select the UDP protocol in Connection protocol Select Datagram Type 1 in Datagram Mode I...

Страница 35: ...address is by using the Hyperterminal program and the serial line of computer The cable s wiring diagram is the classic crossover The configuration of the Hyperterminal must be Baud rate 9600 Bauds Da...

Страница 36: ...4 16 COMMUNICATION PROTOCOLS CHAPTER 4 DC 24 and DC 25 operation manual The attached Hyperterminal set up have been verified and work correctly but any other set up may work also correctly...

Страница 37: ...n the next steps must be followed 1 Connect the serial cable see 4 3 IP address between the computer and the display 2 Open Hyperterminal 3 Select the display s parameter 7 See Error No se encuentra e...

Страница 38: ...4 18 COMMUNICATION PROTOCOLS CHAPTER 4 DC 24 and DC 25 operation manual 6 At this time you have 3 seconds to push the Intro key on your keyboard 7 The following screen is shown...

Страница 39: ...manual 8 You must set up 0 Server Intro 4 WLAN Intro Example of Server Set up Ask your network administrator the IP and Gateway address Example of WLAN Set up Ask your network administrator the corre...

Страница 40: ...s you must first select the equipment by clicking on the XPort 03 to which you wish to assign the address Then press Assign IP and follow the instructions IMPORTANT All units are dispatched from the f...

Страница 41: ...oduct complies with the following European directives Directive LVD 2006 95 CEE Low Voltage Directive Standard UNE EN61010 1 Security in electric equipment Directive EMC 2014 30 UE Electromagnetic Com...

Отзывы: