Spartan 730/730IS Dosimeter Manual
Setting a Visual Alarm on the Spartan 730
40
3.3 Setting a Visual Alarm on the Spartan 730
The Spartan 730’s visual alarm settings enable you and the worker to be aware of
milestones within the current noise exposure period. The Dosimeter 1 setting
comes with 2 customizable visual alarm settings: Alarm 1, Alarm 2, which can be
enabled separately, as shown in section
3.3.1 Enabling Dosimeter 1 Alarm 1
.
First, enable the alarm, then choose the alarm
Source
and
Levels
(noise
exposure metric and levels) to base the alarm on. Each alarm has an action level
and a limit level alarm. These levels are selectable from 70 to 100 dB. The alarm
state is active while the measured sound pressure level is above the set alarm
level. The alarm deactivates when the measured sound pressure level or Dose
level drops below the alarm level. This process is described here in section
.
IN THIS SECTION:
•
3.3.1 Enabling Dosimeter 1 Alarm 1
•
Suggested Application:
•
Set a Limit Level alarm that indicates when the subject experiences an
LCpeak noise exposure.
•
Set an Action Level alarm that indicates when the subject’s noise
exposure reaches 100% Dose.
3.3.1 Enabling Dosimeter 1 Alarm 1
Before you begin:
•
Power on the dosimeter.
•
Stop any measurement in progress.
In this section, use
UP
or
DOWN
on the dosimeter to
highlight an option. Use
SELECT
to set the option. Quick
press
power
to step up one level in a menu.