HVM200 Reference Manual
Mechanical
B-2
B.1.2 ADP081A
FIGURE B-2
(Handle Adapter with SEN04XF Accelerometer)
B.1.3 ADP082A
FIGURE B-3
(Clamp Adapter with SEN04XF Accelerometer)
B.1.4 Measurements
Frequency Response
The frequency response measurements were performed by suspending the test object and exciting it with a
modal hammer. The responses were measured in x, y and z directions using a triaxial accelerometer con-
nected to the test object using the specified adapter. A graphic is included to illustrate the test configuration.
Triaxial Accelerometer
The triaxial accelerometer used for these tests was a Larson Davis Model SEN041F having a sensitivity of 10
mV/g.
Содержание HVM200
Страница 2: ...IHVM200 01 Rev C Larson Davis HVM200 Reference Manual...
Страница 48: ...HVM200 Reference Manual Frequency Weighting Curves A 9...
Страница 49: ...HVM200 Reference Manual Frequency Weighting Curves A 10...
Страница 50: ...HVM200 Reference Manual Frequency Weighting Curves A 11...
Страница 51: ...HVM200 Reference Manual Frequency Weighting Curves A 12...
Страница 63: ...HVM200 Reference Manual Integration Weighting Limits A 24...
Страница 64: ...HVM200 Reference Manual Integration Weighting Limits A 25...
Страница 65: ...HVM200 Reference Manual Integration Weighting Limits A 26...
Страница 66: ...HVM200 Reference Manual Integration Weighting Limits A 27...
Страница 67: ...HVM200 Reference Manual Integration Weighting Limits A 28...