HVM200 Reference Manual
Setting Up the Measurement
3-2
Step 3
Click the Menu icon.
FIGURE 3-1
Menu Icon
TAKE NOTE
A calibration can only
be made if the meter is stopped and
currently on
Live
. Performing a cali-
bration will automatically put the
meter in this state.
Step 4
Click
Cal. Check
on Menu.
FIGURE 3-2
Cal. Check Option
Step 5
Enter the
Reference
value for the shaker you will use to per-
form the calibration check and click
Set
. This value is usu-
ally provided in the shaker reference documentation.
Step 6
Attach the transducer on your shaker so that the axis you
wish to check is oriented properly.
Step 7
Start the shaker and note the
a
rms
values for each axis, as
follows:
TAKE NOTE
The filter is automati-
cally set to the Fb weighting during
calibration check measurements and
restored when finished.
•
Values are gray when level has not been checked.
•
Values are white when level is being checked.
•
Values are red if the axis measurement is complete and the
level varies from the reference value by more than ± 5%.
•
Values are green if the axis measurement is complete and
the level is within ± 5% of the reference value.
Step 8
Repeat the process for each axis on the accelerometer.
Содержание HVM200
Страница 2: ...IHVM200 01 Rev C Larson Davis HVM200 Reference Manual...
Страница 48: ...HVM200 Reference Manual Frequency Weighting Curves A 9...
Страница 49: ...HVM200 Reference Manual Frequency Weighting Curves A 10...
Страница 50: ...HVM200 Reference Manual Frequency Weighting Curves A 11...
Страница 51: ...HVM200 Reference Manual Frequency Weighting Curves A 12...
Страница 63: ...HVM200 Reference Manual Integration Weighting Limits A 24...
Страница 64: ...HVM200 Reference Manual Integration Weighting Limits A 25...
Страница 65: ...HVM200 Reference Manual Integration Weighting Limits A 26...
Страница 66: ...HVM200 Reference Manual Integration Weighting Limits A 27...
Страница 67: ...HVM200 Reference Manual Integration Weighting Limits A 28...