Separation of an RF Cable from the HR Probe 

 
Required tools: 

-

  Open-end wrench 1/4 inch 

-

  Open-end wrench 8 mm (5/16 inch) 

 

Do not use the torque wrench to loosen the connection! 

 
1.  Use the 0.25 inch open-end wrench to prevent the probe from rotating. 
 
2.  Use the 8 mm (5/16 in.) open-end wrench to loosen the connection. 
 
3.  Complete  the  process  by  hand,  turning  only  the  connector  nut  with  your 

fingers. 

 
4.  Carefully pull the connectors apart (do not tilt or twist). 
 

 

Operational Notes 

 

Reproducible  and  reliable  conditions  are  obtained  when  the  probe  tip  is  placed 
vertically and centrally on an RF trace. If the probe tip is placed directly on the RF 
trace a defined distance (about 0.5 mm) between the RF trace and the measuring 
electrode can be achieved. 
 
Resonance processes can be excited in the case of asymmetrical placement and 
with  PCBs  whose  width  is  in  the  range  of  one  centimeter  (the  ground  plane  of 
PCBs which are too small can be increased by connecting them electrically with a 
larger ground plane). 
 
The measuring tip is decoupled from the probe shaft. Touching the probe near the 
probe tip may interfere with the probe function. 
 
The  measuring  cable  must  be  suitable  for  this  frequency  range.  The  measuring 
cable is available on request. 

 