TESTING

Conditioning the Electrode

The TRACER electrode comes in a factory cleaned condition. If the meter is new 

or has been in storage for an extended period of time, at least two tests should be 

performed to condition the electrode. The electrode will perform best when in 

continuous use. Gently wipe the electrode surface with a paper towel between uses.

Measurement

1. Fill a sample cup with exactly 20 ml of the sample water.

NOTE: Turbidity of the solution has no effect on the reading.

2. Add one TRACER TCL Tablet (7044A).
3. Use the tablet crusher (0175) to crush the tablet.
4. Cap and shake vigorously for 20-30 seconds. The tablet should be

completely disintegrated.

5. Turn the TRACER on. Wait 3 seconds for self-calibration. Immediately

immerse in the sample.

6. Stir the sample with the TRACER for 10 seconds. Stop stirring.

The sample and the electrode must remain still and stable during

the entire measurement.

7. The display will f ash while readings are changing rapidly. The result

should be read when the display stabilizes. The reading will be held and

the HOLD annunciator will come on after 120 seconds.

8. Remove the TRACER from the sample. Turn the unit off. Replace

electrode cap.

NOTE: If a series of subsequent measurements is to be carried out, brief y

rinse the electrode with water (deionized, if available) and shake off.

Gently wipe the electrode surface with a paper towel between uses.

9. The TRACER must be turned off and on again between samples.

Storing Readings

1. When the display is steady and not f ashing, press the MODE/HOLD button

to store the data. The storage location number, the stored reading and HOLD 

will be displayed. Press MODE/HOLD to return to the current reading.

2. Only one reading can be stored during the 120 second measurement cycle.
3. After f fteen readings are stored, the sixteenth stored reading will over-write

the f rst stored reading.

Содержание Tracer Total Chlorine PockeTester 1740

Страница 1: ...Water Testing Leader Since 1919 Tracer Total Chlorine PockeTesterTM Code 1740 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ...WARNING This set contains chemicals that may be harmful if misued Read cautions on individual containers carefully Not to be used by children except under adult supervision...

Страница 3: ...ging the Display Temperature Units 6 Low Battery Indicator 6 Testing Conditioning the Electrode 7 Measurement 7 Storing Readings 7 Recalling Stored Readings 8 For Best Accuracy 8 Offset Adjustment Pro...

Страница 4: ...buffer be added to the sample before testing SPECIFICATIONS Display Multifunction display with bar graph Operating conditions 32 to 122 F 0 to 50 C and 80 RH Chlorine range 0 01 to 10 00 ppm Total Ch...

Страница 5: ...compartment cover 2 LCD Display 3 MODE HOLD button 4 CAL RECALL button 5 ON OFF button 6 Electrode Retaining Collar ring 7 Electrode Sensor Note The Electrode storage cap is not shown LCD Display 1 B...

Страница 6: ...R is turned on it will automatically self calibrate and the SELF and CAL icons will be displayed These icons will disappear after the self calibration is complete The main display and bar graph will t...

Страница 7: ...and the electrode must remain still and stable during the entire measurement 7 The display will f ash while readings are changing rapidly The result should be read when the display stabilizes The read...

Страница 8: ...use 3 If measuring solutions with large differences in chlorine concentration for example 0 1 ppm and 5 0 ppm use a separate sample cup for each concentration 4 Discard the sample cup when it becomes...

Страница 9: ...oximately 5 seconds until C_O appears at the bottom of the display 5 IMMEDIATELY press the MODE HOLD button repeatedly to increase the value or press the CAL RECALL button repeatedly to decrease the v...

Страница 10: ...1 00 mL of the 10 000 ppm standard into another 100 mL volumetric f ask Dilute to the 100 mL line with deionized water Cap and mix This standard is 100 ppm 3 Pipet exactly 1 00 mL of the 100 ppm solu...

Страница 11: ...splayed keep the electrode in standard and press the CAL key again until CAL is displayed then press the CAL and MODE keys again simultaneously Immediately release both keys 9 After approximately 5 to...

Страница 12: ...tabs 3 Replace the four CR2032 batteries Observe polarity 4 Replace the battery compartment cap Cleaning the Electrode The TRACER Chlorine Electrode arrives in a factory cleaned condition and is read...

Страница 13: ...meter to malfunction 5 Firmly tighten the electrode collar to create a seal with the rubber gasket between the electrode and the meter Do not over tighten 6 Cover the connector end of the removed ele...

Страница 14: ...ment to be free of defects in parts and workmanship for 1 year from the date of shipment and the probe to be free of defects in parts and workmanship for 6 months from the date of shipment If it shoul...

Отзывы: