Replacing the Electrode

The TRACER is shipped with an electrode attached. If the electrode needs to be 

replaced or changed, follow these steps.
1. Press the ON/OFF button to turn the meter off.
2. To remove an electrode, unscrew and completely remove the electrode

collar. Turn the collar counter-clockwise.

3. Gently rock the electrode from side to side, pulling it downwards, until it

disconnects from the meter.

4. To attach an electrode, align the slots and carefully plug the electrode into

the meter socket. CAUTION: Take care to align pins carefully. Bent or

broken pins will cause the meter to malfunction.

5. Firmly tighten the electrode collar to create a seal with the rubber gasket

between the electrode and the meter. Do not over tighten.

6. Cover the connector end of the removed electrode with the clear plastic cap.

TROUBLESHOOTING

Problem

Check

Action

Power on but no display Batteries

Insert batteries
Verify correct polarity
Replace

Unstable reading

Electrode

Immerse probe more deeply 

in sample
Remove air bubbles caught 

under electrode
Clean electrode
Replace electrode

Slow response

Electrode

Clean electrode

Unexpected low results

Tablet

Conf rm tablet has been added

Unexpected high results Tablet

Dissolve tablet completely

Display frozen

HOLD function

Press MODE/HOLD or turn 

meter off.Restart.

Button press

Remove batteries (stored data 

will be lost)

Содержание Tracer Total Chlorine PockeTester 1740

Страница 1: ...Water Testing Leader Since 1919 Tracer Total Chlorine PockeTesterTM Code 1740 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ...WARNING This set contains chemicals that may be harmful if misued Read cautions on individual containers carefully Not to be used by children except under adult supervision...

Страница 3: ...ging the Display Temperature Units 6 Low Battery Indicator 6 Testing Conditioning the Electrode 7 Measurement 7 Storing Readings 7 Recalling Stored Readings 8 For Best Accuracy 8 Offset Adjustment Pro...

Страница 4: ...buffer be added to the sample before testing SPECIFICATIONS Display Multifunction display with bar graph Operating conditions 32 to 122 F 0 to 50 C and 80 RH Chlorine range 0 01 to 10 00 ppm Total Ch...

Страница 5: ...compartment cover 2 LCD Display 3 MODE HOLD button 4 CAL RECALL button 5 ON OFF button 6 Electrode Retaining Collar ring 7 Electrode Sensor Note The Electrode storage cap is not shown LCD Display 1 B...

Страница 6: ...R is turned on it will automatically self calibrate and the SELF and CAL icons will be displayed These icons will disappear after the self calibration is complete The main display and bar graph will t...

Страница 7: ...and the electrode must remain still and stable during the entire measurement 7 The display will f ash while readings are changing rapidly The result should be read when the display stabilizes The read...

Страница 8: ...use 3 If measuring solutions with large differences in chlorine concentration for example 0 1 ppm and 5 0 ppm use a separate sample cup for each concentration 4 Discard the sample cup when it becomes...

Страница 9: ...oximately 5 seconds until C_O appears at the bottom of the display 5 IMMEDIATELY press the MODE HOLD button repeatedly to increase the value or press the CAL RECALL button repeatedly to decrease the v...

Страница 10: ...1 00 mL of the 10 000 ppm standard into another 100 mL volumetric f ask Dilute to the 100 mL line with deionized water Cap and mix This standard is 100 ppm 3 Pipet exactly 1 00 mL of the 100 ppm solu...

Страница 11: ...splayed keep the electrode in standard and press the CAL key again until CAL is displayed then press the CAL and MODE keys again simultaneously Immediately release both keys 9 After approximately 5 to...

Страница 12: ...tabs 3 Replace the four CR2032 batteries Observe polarity 4 Replace the battery compartment cap Cleaning the Electrode The TRACER Chlorine Electrode arrives in a factory cleaned condition and is read...

Страница 13: ...meter to malfunction 5 Firmly tighten the electrode collar to create a seal with the rubber gasket between the electrode and the meter Do not over tighten 6 Cover the connector end of the removed ele...

Страница 14: ...ment to be free of defects in parts and workmanship for 1 year from the date of shipment and the probe to be free of defects in parts and workmanship for 6 months from the date of shipment If it shoul...

Отзывы: