13
OPERATIONAL MATRIX
Function /
Resulting
Action
Power
Status
Mode
Setting
Required Button
Press Sequence
Comments
ON/OFF
Any
Any
Momentary press of
the
ON/OFF
button
Water Saturated
Air Calibration
On
Any
Place electrode in
electrode cap. Press
& hold
CAL
button
for 2 seconds.
Zero
Calibration
On
Any
Place electrode in
zero DO solution.
Wait for stability.
Press & hold
CAL
button for 2 seconds.
Also works
with probe
removed.
(Cal at zero
current)
Store Reading
On
Any
Momentary press
of
MODE
button.
Stores and
holds reading.
‘HOLD’
displayed
Hold Release
On
While In
Hold Mode
Momentary press of
the
MODE
button.
Enter Memory
Retrieval
On
Any
Momentary press
of
CAL
button
followed by a
momentary press of
the
MODE
button
within 4 seconds.
If no data
is stored in
memory, ‘End’
is displayed
brief y, and
then returns to
last mode.
Scroll Stored
Readings
On
Memory
Recall
Momentary press
of
MODE
button.
Exit Memory
Retrieval
On
Memory
Recall
Momentary press
of
CAL
button.
Clear Stored
Memory
On
Any
Measure-
ment Mode
Press and hold
ON/OFF
button
for 4 seconds
‘clr’ is
displayed.