![LaMotte Tracer DO PockeTester Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/lamotte/tracer-do-pocketester/tracer-do-pocketester_manual_3430965008.webp)
8
Auto-Power Off Feature
The auto power off feature automatically shuts the meter off 10 minutes after the
most recent button press. To disable this feature, refer to the Disable Auto-Power
Off section.
Disable Auto-Power Off Feature
With the meter
ON
, momentarily press
CAL
, then quickly press and hold both
MODE
and
ON/OFF
until ‘oFF’ is displayed. To restore the Auto Power Off
Feature turn the meter off and on again using
ON/OFF
.
Low Battery Indicator
When the battery voltage falls below the operating threshold, ‘BAT’ will
appear on the display. Refer to the Maintenance section for battery replacement
information.
Storing Readings
1. Press
MODE
to store a reading. The storage location number will be shown
on the lower display and the main display will show the stored reading.
The meter will enter the HOLD mode and ‘HOLD’ will appear on the display.
2. Press
MODE
again to exit the HOLD mode and return to normal operation.
The next time
MODE
is momentarily pressed another reading will be
stored and so on.
3. If more than 25 readings are stored, previously stored readings, starting
with reading number 1, will be overwritten.
Recalling Stored Readings
1. Momentarily press the
CAL
and then, within 4 seconds, momentarily press
MODE
. The last stored data point location will be displayed (1 to 25).
Each time
MODE
is momentarily pressed the next most recently stored
data point will be displayed.
2. After the last data point stored is displayed, press
MODE
again to return
the display to the beginning of the list.
3. Press
CAL
at any time to stop the data retrieval process and return the meter
to the normal measurement mode.
Clearing Stored Readings
1. With the meter ON press and hold
ON/OFF
for 4 seconds.
2. When ‘clr’ is shown on the display the memory is cleared.