5.
Plot results on graph paper or computer using any available plotting
program. If results are as %T versus concentration, semilog graph paper
must be used. Plot the standard solution concentrations on the horizontal,
linear axis, and the %T on the vertical, logarithmic axis. If results are as
absorbance versus standard solution concentration, simple linear graph
paper can be used. Plot the standard solution concentration on the
horizontal axis, and the absorbance on the vertical axis.
6.
After plotting the results, draw a line, or curve, of best fit through the
plotted points. The best fit may not connect the points. There should be
approximately an equal number of points above the curve as below the
curve. Some reagent systems will produce a straight line, while others
produce a curve. Many computer spreadsheet programs can produce the
curve of best fit by regression analysis of the standard solution data.
A sample of each type of graph appears below:
SMART SPECTRO SPECTROPHOTOMETER 05.04
15
0
1
10
100
1 2 3 4 5 6
Concentration in ppm
%T vs. Concentration
CALIBRATION CURVE
%T
Tr
ansmission
8 9 10
7
0
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
1 2 3 4 5 6
Concentration in ppm
Absorbance vs. Concentration
CALIBRATION CURVE
Absorbance
8 9 10
7