4.  Scroll to and select

 045 Fluoride 

from menu.

045 Fluoride

Scan Blank

Scan Sample

12:00:00                       001/500                       

5.  This test requires a reagent blank. 

Rinse a clean colorimeter tube 

(0290) with clear, colorless, Fluoride 

free water. Fill to the 10 mL line with 

clear, colorless, Fluoride free water.

6.  Use the 0.5 mL pipet (0353) to add 

0.5 mL of *Sodium Arsenite Solution 

(4128). Cap and mix.

7.  Use the 1.0 mL pipet (0354) to 

add 2 measures of *Acid-Zirconyl 

SPADNS Reagent (3875). Cap and 

mix thoroughly. (This is the reagent 

blank.) Dry the tube with a lint-free 

cloth.

8.  Insert tube into chamber, close lid 

and select 

Scan Blank

.

9.  Rinse a clean colorimeter tube 

(0290) with sample water. Fill to the 

10 mL line with sample water.

Содержание DC1500 FLUORIDE

Страница 1: ...DC1500 FLUORIDE Fluoride Colorimeter Kit Code 3243 ACCEPTED...

Страница 2: ...r Connection Output 15 Computer Connection 15 WaterLink Connect 2 15 Battery Battery AC Operation 15 Battery Replacement 16 Maintenance Cleaning 16 Returns 16 Meter Disposal 16 General Operating Infor...

Страница 3: ...1720 01 that is supplied with the kit Make no substitutions Test Tubes with Caps 0290 6 Replacement Chamber 3 0038 USB Cable 1720 01 USB Wall Adapter 1721 Car Charger 5 0132 TEST METHODS SPECIFICATIO...

Страница 4: ...the concentration given Alkalinity CaCO3 5000 mg L Aluminum Al 3 0 1 mg L Iron Fe 3 10 mg L Sulfate SO4 2 200 mg L Color and turbidity must be removed or compensated for in the procedure Temperature s...

Страница 5: ...ear colorless Fluoride free water 6 Use the 0 5 mL pipet 0353 to add 0 5 mL of Sodium Arsenite Solution 4128 Cap and mix 7 Use the 1 0 mL pipet 0354 to add 2 measures of Acid Zirconyl SPADNS Reagent 3...

Страница 6: ...t blank NOTE The meter will remember the last scanned blank reading It is not necessary to scan a blank each time the test is performed To use the previous blank reading instead of scanning a new one...

Страница 7: ...select TestingMenu All Tests 045 Fluoride 12 00 00 001 500 3 Scroll to and select 045 Fluoride from menu 045 Fluoride Scan Blank Scan Sample 12 00 00 001 500 4 This test requires a reagent blank Rins...

Страница 8: ...imeter tube 0290 with sample water Fill to the 10 mL line with sample water 9 Use the 0 5 mL pipet 0353 to add 0 5 mL of Sodium Arsenite Solution 4128 Cap and mix 10 Use the 1 0 mL pipet 0354 to add 2...

Страница 9: ...he value shown to the concentration of the prepared standard 1 0 in this example NOTE A maximum adjustment of 25 is possible 045 Fluoride 1 0 ppm Edit ENTER Save ENTER Default 12 00 00 001 500 15 Pres...

Страница 10: ...ach Set Time Year 2000 12 00 00 001 500 3 Press ENTER to select the final character The time and date will be saved and the meter will return to the Editing Menu Editing Menu Set Clock Logging Factory...

Страница 11: ...Editing Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 FACTORY SETUP The Factory Setup menu is used in manufacturing of the colorimeter This menu is not for use by the operator i...

Страница 12: ...iewing The default setting is 10 seconds If Button Control is chosen the backlight button on the key pad will act as an on off switch and the backlight will remain on or off when the meter is being us...

Страница 13: ...d the meter will return to the Editing Menu Editing Menu Logging Factory Setup Set Power Save Set Backlight Time 12 00 00 001 500 SELECTING A LANGUAGE There are eight languages available English Spani...

Страница 14: ...incorporate a looping feature which allows the user to quickly reach the last choice in the menu from the first choice In a looping menu the last choices in the menu are above the first choice and scr...

Страница 15: ...B mini B connector into the meter and the larger end of the USB cable USB type A connector into the wall adapter Plug the wall adapter into an AC outlet Reinsert the rubber USB port plug after chargin...

Страница 16: ...the backlight time option to 10 seconds or select Button Control and keep the backlight off BATTERY REPLACEMENT The lithium ion battery used in this unit should last for many years with normal use Whe...

Страница 17: ...with an integrated interference filter The interference filter selects a narrow band of light from the corresponding LED for the colorimetric measurements A USB wall adapter USB computer connection o...

Страница 18: ...t the top of the menu will lead to the bottom of the menu Header Menu Title Main Window Selection First Choice Second Choice Third Choice Another Footer 12 00 00 001 500 P And Another And So On A ligh...

Страница 19: ...prevent dust from entering the tubes Dry tubes should be stored with the caps on to prevent contamination After a tube has been filled and capped it should be held by the cap and the outside surface s...

Страница 20: ...mpany to receive compensation for damaged goods GENERAL PRECAUTIONS READTHEINSTRUCTIONMANUALBEFOREATTEMPTINGTOSETUPOROPERATETHE METER Failure to do so could result in personal injury or damage to the...

Страница 21: ...RH non condensing USB Port Mini B Power Requirements USB wall adapter USB computer connection or lithium ion rechargeable battery Battery Charge Life Approximately 380 tests with backlight on to 1000...

Страница 22: ...no other warranty whether written or oral is expressed or implied TROUBLESHOOTING ERROR MESSAGES Over Range If the message OVERRANGE is displayed when scanning a sample the sample may be over range o...

Страница 23: ...B wall computer adapter Shut Down Low Batt Shut Down Battery is too low to operate the unit Charge battery or use USB wall computer adapter Overrange Sample is outside of acceptable range Dilute sampl...

Страница 24: ...gton Ave Chestertown Maryland 21620 USA 1 410 778 3100 800 344 3100 www lamotte com LaMotte and WaterLink are registered trademarks of LaMotte Company 2023 LaMotte Company All Rights Reserved 3243 MN...

Отзывы: