9
12. Press
to scroll to
Calibrate
.
045 Fluoride
0.9
ppm
%T/Abs
Calibrate
12:00:00 001/500
13. Press
ENTER
to select
Calibrate
.
A reverse font (light background
with dark characters) will appear
to indicate that the reading can be
adjusted.
045 Fluoride
0.9
ppm
∧
,
∨
=Edit, ENTER=Save
∧
+ENTER=Default
12:00:00 001/500
14. Press
or
to adjust the
value shown to the concentration
of the prepared standard, 1.0 in this
example.
NOTE: A maximum adjustment of
25% is possible.
045 Fluoride
1.0
ppm
∧
,
∨
=Edit, ENTER=Save
∧
+ENTER=Default
12:00:00 001/500
15. Press
ENTER
to save the value.
To leave the Calibration procedure
without saving the adjustment,
press
EXIT
.
Press
and
ENTER
at any time
to return to the default value.
The calibration has now been
standardized and can be used for
testing. Scroll to
Scan Blank
and
begin testing.
045 Fluoride
1.0
ppm
%T/Abs
Calibrate
12:00:00 001/500