SAMPLE DILUTION TECHNIQUES

If a test result is out of the range of the meter, it must be diluted. The test 
should then be repeated on the diluted sample. The following table gives quick 
reference guidelines for dilutions of various proportions.

Amount of Sample

Deionized Water to Bring 
Final Volume to 10 mL

Multiplication Factor

10 mL

0 mL

1

5 mL

5 mL

2

2.5 mL

7.5 mL

4

1 mL

9 mL

10

0.5 mL

9.5 mL

20

All dilutions are based on a fi nal volume of 10 mL, so several dilutions will 
require small volumes of the water sample. Graduated pipets should be 
used for all dilutions. If volumetric glassware is not available, dilutions can 
be made with the colorimeter tube. Fill the tube to the 10 mL line with the 
sample and then transfer it to another container. Add 10 mL volumes of 
deionized water to the container and mix. Transfer 10 mL of the diluted 
sample to the colorimeter tube and follow the test procedure. Repeat the 
dilution and testing procedures until the result falls within the range of the 
calibration. Multiply the test result by the dilution factor. For example, if 10 mL 
of the sample water is diluted with three 10 mL volumes of deionized water, 
the dilution factor is four. The test result of the diluted sample should be 
multiplied by four.

GENERAL INFORMATION

 

PACKAGING AND DELIVERY

Experienced packaging personnel at LaMotte Company assure adequate 
protection against normal hazards encountered in transportation of 
shipments.

After the product leaves LaMotte Company, all responsibility for safe delivery 
is assured by the transportation company. Damage claims must be fi led 
immediately with the transportation company to receive compensation for 
damaged goods.

 

GENERAL PRECAUTIONS

READ THE INSTRUCTION MANUAL BEFORE ATTEMPTING TO SET UP OR 
OPERATE THE METER. 

Failure to do so could result in personal injury or 

damage to the meter. The meter should not be used or stored in a wet or 
corrosive environment. Care should be taken to prevent water from wet tubes 
from entering the meter chamber.

NEVER PUT WET TUBES IN THE METER.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание DC1500 Copper

Страница 1: ...DC1500 COPPER Copper Colorimeter Kit Code 3245 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...ting Power Save 14 n Setting Backlight Time 15 n Selecting A Language 16 n Looping Menus 17 Computer Connection n Output 18 n Computer Connection 18 Battery n Battery AC Operation 18 n Battery Replace...

Страница 3: ...s 25 n Limits of Liability 25 n Speci cations 26 n CE 27 n IP67 27 n Warranty 27 n Register Your Meter 27 Trouble Shooting Guide n Error Messages 28 n Calibration 28 n Stray Light 28 n Troubleshooting...

Страница 4: ...r digit reagent code number listed on the reagent label in the contents list or in the test procedures Omit any letter that follows or precedes the four digit code number For example if code is 4450WT...

Страница 5: ...ults in the addition of copper into the water supply APPLICATION Drinking surface and saline waters domestic and industrial wastes RANGE 0 0 to 6 0 ppm Copper METHOD Cupric ions form a yellow colored...

Страница 6: ...ress ENTER to select Testing Menu All Tests 036 Cu Thiocarbamate 12 00 00 001 500 4 Scroll to and select 036 Cu Thiocarbamate from menu 036 Cu Thiocarbamate Scan Blank Scan Sample 12 00 00 001 500 5 R...

Страница 7: ...h time the test is performed To use the previous blank reading instead of scanning a new one scroll to Scan Sample and proceed For the most accurate results the meter should be blanked before each tes...

Страница 8: ...he LaMotte logo screen will appear for about 3 seconds and the Main Menu will appear Main Menu Testing Menu Editing Menu 12 00 00 001 500 2 Press ENTER to select Testing Menu All Tests 036 Cu Thiocarb...

Страница 9: ...to mix Wipe tube dry 8 Insert tube into chamber Close lid Select Scan Sample 036 Cu Thiocarbamate 0 99 ppm Scan BLank Scan Sample 12 00 00 001 500 9 Press to scroll to Calibrate 036 Cu Thiocarbamate...

Страница 10: ...m adjustment of 25 is possible 036 Cu Thiocarbamate 1 00 ppm Edit ENTER Save ENTER Default 12 00 00 001 500 12 Press ENTER to save the value To leave the Calibration procedure without saving the adjus...

Страница 11: ...year is displayed Press or to scroll to the appropriate character Press ENTER to select the character The month day hour format hour minute second AM PM will be dislpayed Repeat for each Set Time Yea...

Страница 12: ...diting Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 2 Press ENTER to select Logging Logging Display Test Log Logging Enabled Logging Disabled Erase Log 12 00 00 001 500 3 Press...

Страница 13: ...er Save Editing Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 2 Press ENTER to select Set Power Save Set Power Save Disable 5 Minutes 15 Minutes 30 Minutes 12 00 00 001 500 3 Pr...

Страница 14: ...attery will have to be charged if the USB Wall Adapter is not being used 1 From the Editing Menu press or to scroll to Backlight Time Editing Menu Logging Factory Setup Set Power Save Set Backlight Ti...

Страница 15: ...anguage Select Language English Spanish French Portugese 12 00 00 001 500 4 Press ENTER The screen will display Storing for about 1 second and the meter will return to the Editing Menu Editing Menu Se...

Страница 16: ...ow demonstrate a looping menu AND SO ON AND SO ON AND SO ON THIRD TO LAST LAST CHOICE LAST CHOICE SECOND TO LAST AND SO ON AND SO ON LAST CHOICE TESTING MENU TESTING MENU TESTING MENU FIRST CHOICE SEC...

Страница 17: ...USB cable USB mini B connector into the meter and the larger end of the USB cable USB Type A connector into a USB port on a computer The battery icon will show no bars and ash when the unit rst turns...

Страница 18: ...backlight o BATTERY REPLACEMENT The lithium ion battery used in this unit should last for many years with normal use When it no longer powers the unit long enough to meet testing requirements it will...

Страница 19: ...he user to select options from the menu driven software to directly read test results or to review stored results of previous tests in the data logger The menus can be displayed in seven di erent lang...

Страница 20: ...isplays the title of the current menu The footer line displays the time and the date the data logger status and the battery status The menu selection window is in the middle of the display between the...

Страница 21: ...tte logo is displayed The top two lines of the display are used to display information in a large easy to read format The menus work in the same way as previously described but two lines of the menu a...

Страница 22: ...n an inverted position to prevent dust from entering the tubes Dry tubes should be stored with the caps on to prevent contamination After a tube has been lled and capped it should be held by the cap a...

Страница 23: ...e result falls within the range of the calibration Multiply the test result by the dilution factor For example if 10 mL of the sample water is diluted with three 10 mL volumes of deionized water the d...

Страница 24: ...nd at Keep equipment and reagent chemicals out of the reach of young children Ensure that the protection provided by this equipment is not impaired Do not intall or use this equipment in a manner that...

Страница 25: ...lection Languages English Spanish French Portuguese Italian Chinese Japanese Temperature Operation 0 50 C Storage 40 60 C Operation Humidity Range 0 90 RH non condensing USB Port Mini B Power Requirem...

Страница 26: ...ctromagnetic compatibility and safety To view certi cates of compliance go to IP67 Certi cation The meets IP67 standards for protection against dust and immersion only when the USB port plug is in pla...

Страница 27: ...00 00 001 500 CALIBRATION As with all pre calibrated meters it is highly recommended even if not required by regulations that the user periodically verify the performance of the meter by running stand...

Страница 28: ...Shut Down Battery is too low to operate the unit Charge battery or use USB wall computer adapter Overrange Sample is outside of acceptable range Dilute sample and test again Unusually large negative...

Отзывы: