6. TROUBLESHOOTING
Problem
Cause
Solution
No display
Batteries not in place.
a) Insert batteries.
b) Re-insert batteries in
correct polarity.
“---“ on
display
Meter does not have 2
point calibration
Perform 2 or 3 point ion
calibration.
“LO” displays
in the
DISPLAY
Low battery
Replace batteries.
Unstable
reading
a) Electrode not deep
enough in sample
b) Dirty electrode.
c) Broken electrode
a) Place electrode deeper
in sample.
b) Clean electrode and
recalibrate.
c) Replace electrode.
“Er1” display
Bu
ff
er value out of
tolerance
Use fresh pH bu
ff
er
solution and recalibrate.
Ensure correct pH bu
ff
er
group was selected.
“Er2” display
Single point calibration
Perform at least 2 point
calibration.
“Er3” display
ISE slope not within the
speci
fi
ed tolerance
Check ISE is in good
working condition-refer to
ISE section of manual.
“Er4” display
Calibration points not
within 1 decade
Ensure calibration points
are within 1 decade of
each other.
Not able to
calibrate
a) Display freezes
b) Faulty electrode
c) Inaccurate bu
ff
er
a) Release reading by
pressing HOLD.
b) Replace electrode.
c) Replace expired bu
ff
er
solutions.