16
3. Press the
CAL
key to enter the temperature calibration mode.
The DISPLAY will show “
CA
” momentarily and the displayed
reading will
fl
ash.
4. Press the
or
keys until the display shows the desired
temperature. The meter will allow an adjustable maximum value
of ± 5 °C from the factory default.
5. To cancel or abort this operation, press the
CAL
key. To con
fi
rm
the calibration, press the
ENTER
key. A new value will be
stored in the meter non-volatile memory. The meter will display
“
CO
” momentarily, and then revert to the measurement mode.
3.3.2 Without Temperature Probe (no ATC)
If a temperature probe is not used, the meter will compensate the pH
response based on a temperature value that has been manually set or at
the factory default of 25.0 °C.
1. Press the
MODE
key until “
°C
” shows on the DISPLAY.
2. Compare the displayed value to a NIST certi
fi
ed thermometer or
thermometer that is known to be accurate (dipped into a constant
temperature bath).
3. Press the
CAL
key to enter the temperature calibration mode.
The DISPLAY will show “
CA
” momentarily and the displayed
reading will
fl
ash. Note that the displayed value should be either
25.0 °C or the last set temperature value.
4. Press the
or
keys until the displays shows the desired
temperature. The value can be set from 0 to 100 °C.