13

3.1.3 Resetting User Calibrated Values

The calibrated pH/mV/Ion values can be reset to the factory default using 

the procedure below.  The temperature o

 set will not be reset using this 

procedure. To abort the reset, press 

CAL

 to revert to the measurement 

mode. 

1.  Press and hold 

CAL

 while switching the meter on using the 

ON/

OFF

 key. The DISPLAY will show “

rSt

” blinking. 

2. Press the 

ENTER

 key to con

fi

 rm. The meter will automatically 

clear all stored pH/Ion calibrations or mV o

 set values and 

revert to the measurement mode. 

3.2 Ion Calibration

The meter is capable of a 2 or 3 point ion calibration with standard 

solutions. The meter will display “- - -“ if the meter has not been 

calibrated or if the meter has been reset.
To abort the ion calibration, press

 CAL

 to revert to the measurement 

mode. 
Ion calibration values are not stored in the meter’s non-volatile memory. 

Ion calibration data is lost once the meter is reset and when the batteries 

are removed and replaced.
The error message “

Er2

” will be displayed after a single point calibration 

is completed. Recalibrate using a minimum of 2 points.
Calibration values are successfully stored if the ISE (Ion Selective 

Electrode) slope is within the speci

fi

 ed tolerance of 15-90mV/decade, 

otherwise an error message “

Er3

” will be displayed.

If any of the calibration points are not within 1 decade of each other, an 

error message “

Er4

” will be shown at the end of calibration process. 

The ion calibration options available include 0.1, 1.0, 10.0, 100.0 ppm. 

Recalibrate and ensure that consecutive calibration points are 1 decade 

apart from each other.
Ensure that new or fresh standard solutions are used during the 

calibration. Do not reuse ion standard solutions because they may be 

contaminated and a

 ect the calibration and accuracy of measurements. 

Store standard solutions in a dry, cool environment. Check that electrodes 

and ion standard solutions are kept in good condition for best results. 
Before use, remove the plastic protective cap at the tip of sensor of 

the electrode and refer to the electrode instruction manual for proper 

operation. Rinse the electrode before and after each calibration or sample 

measurement to avoid cross-contamination. 

Содержание 5-1936-01

Страница 1: ...Code 5 1936 01...

Страница 2: ......

Страница 3: ...not represent a commitment on the part of LaMotte Company WARNING Reagents marked with an are considered to be potential health hazards To view or print a Safety Data Sheet SDS for these reagents go t...

Страница 4: ...alibration 10 3 1 1 pH Calibration Procedure 11 3 1 2 Changing the pH Bu er Group 12 3 1 3 Resetting User Calibrated Values 12 3 2 Ion Calibration 13 3 2 1 Ion Calibration Procedure 13 3 3 Temperature...

Страница 5: ...mV ORP temperature and allows direct ion concentration measurement of various ions mono and divalent The mV mode is also useful for diagnosing problems with ion selective electrodes ISE The meter incl...

Страница 6: ...up in the mode that was last used CAL Allows calibration for Ion pH mV or Temperature or to abort calibration and return to measure without confirming a value MODE Selects measurement mode for Ion mV...

Страница 7: ...2 3 Inserting Removing the Rubber Boot Stand 1 To remove the meter from the rubber boot push out the bottom edge of the meter until it is completely out of the boot Ensure that the probe cables are no...

Страница 8: ...he direction of the arrow and lift up Note the polarity of the batteries before inserting them into position After the batteries have been replaced reposition the cover and press down until it locks 2...

Страница 9: ...without temperature compensation 2 7 Conditioning the pH Probe For best results condition the pH probe before the first use or if it has not been used for a long time by soaking it in pH 4 bu er solut...

Страница 10: ...ts perform at least a 2 point calibration at room temperature 25 C using standard bu ers Begin with pH 7 00 USA group pH 6 86 NIST group or pH 6 97 Pb group For a 1 point calibration the calibration s...

Страница 11: ...tandard solution e g pH 7 00 into a clean dry container Dip the pH probe and temperature probe into the solution Stir gently and wait for the reading to stabilize approximately 30 seconds depending on...

Страница 12: ...p selection mode 3 Press MODE to toggle between the available pH bu er groups 4 Press HOLD ENTER to confirm the pH bu er group Note if Pb is selected measurement mode will begin If USA or nSt is selec...

Страница 13: ...e displayed after a single point calibration is completed Recalibrate using a minimum of 2 points Calibration values are successfully stored if the ISE Ion Selective Electrode slope is within the spec...

Страница 14: ...Adjustment solution if required Stir the solution gently with the electrode Press the CAL key to begin the calibration mode 2 The display will show CA to indicate the calibration mode momentarily fol...

Страница 15: ...easurement mode If the calibration is not successfully stored in the memory an error message Er3 will be displayed This occurs when the slope is lower than 15mV decade or higher than 90mV decade 3 3 T...

Страница 16: ...2 Without Temperature Probe no ATC If a temperature probe is not used the meter will compensate the pH response based on a temperature value that has been manually set or at the factory default of 25...

Страница 17: ...erature 3 Dip the probes in the aqueous test sample Stir gently with the probes and wait for the reading to stabilize Note the reading Freeze the displayed if desired for details refer to Section 4 3...

Страница 18: ...results store the ISE dry and the pH ORP probe bulb wet Store the pH ORP glass bulb in pH probe storage solution NEVER use deionized water for storage Wash probes with clean water after each use ISEs...

Страница 19: ...alibrate c Replace electrode Er1 display Bu er value out of tolerance Use fresh pH bu er solution and recalibrate Ensure correct pH bu er group was selected Er2 display Single point calibration Perfor...

Страница 20: ...4 01 7 00 10 01 USA pH 4 01 6 86 9 18 NIST pH 4 10 6 97 Pb Temperature Range 0 0 to 100 0 C Resolution 0 1 C Accuracy 0 5 C Temperature Comp Automatic Manual 0 to 100 C Millivolt Range 500 to 500 mV R...

Страница 21: ...ablets pH 4 0 50 tablets 3983A H Mini Bu er Tablets pH 7 0 50 tablets 3984A H Mini Bu er Tablets pH 10 0 50 tablets 3985A H Ion Selective Electrode Ammonia ISE 5 0043 Fluoride ISE 5 0048 Nitrate ISE 5...

Страница 22: ...pecific purpose and will not be liable for any direct indirect incidental or consequential damages LaMotte Company s total liability is limited to repair or replacement of the product The warranty set...

Страница 23: ...23...

Страница 24: ...PO Box 329 Chestertown MD 21620 800 344 3100 410 778 3100 f 410 778 6394 www lamotte com 626094 30 11 19...

Отзывы: