3.4 Wiring For 24 VDC and 110 VAC Power
Access to all User wiring is through the ¼ DIN junction box in the lower
left on the mounting panel.
The standard on-line turbidimeter requires 24 VDC for power and typically
draws 270 mA (6.5 watts) or optionally requires 100-240 VAC at 50/60 Hz
for power and typically draws up to 250 mA (10 watts).
Pg11 and Pg16 punch out conduit holes are provided for user terminal
connections. We supply Pg11 cord grips that can be removed.
Note: Loop 2 and Relay options are not currently available on 2020
ClearTrace System.
11
Turbidity User
Junction Box
Lamp assembly
screws into sample
chamber lid
Lamp cable connection
Terminal Connection to AV38
USER Terminal Connections:
See next page for wiring tables
24 VDC
or
110 VAC
Lamp
24
VDC
or
110
VA
C
Содержание 2020 ClearTrace Series
Страница 1: ...2020 ClearTrace In line Turbidity System Code 5 0113 24V Code 5 0113 110V...
Страница 2: ......
Страница 35: ......