From the measure screen,
press MENU.
From the Main Menu, use the
up/down arrows to select
Calibrate. Then press
ENTER.
From the Calibrate Menu, use
the up/down arrows to select
1-Point Sample. Then press
ENTER.
The current sensor value is
dynamically displayed and
asks the user to be sure the
measurement is ready for
calibration before proceeding.
When ready, press the
ENTER key.
The calibration value can be
adjusted with the arrow keys.
Use the up/down arrows for
fine adjust and the left/right
arrows for course adjust.
After the calibration value is
adjusted to the desired value,
press the ENTER key.
NOTE: The display may
default to 40 MTU. Adjust to
20.00 NTU.