2. INSTALLATION
2
INSTALLATION
2.1
LINE VOLTAGE
The CA-1550-CML can operate on either 115 or 230 volts AC. The desired line
voltage can be changed by opening the unit and modifying the voltage-set jumper(s).
The unit is factory-preset at 115 volts AC.
WARNING: Ensure that the unit is turned off and that no
battery is connected before attempting to open the cover.
CAUTION: Ensure that the unit is set for the appropriate
line voltage before operation.
A.
Turn off the unit and remove any battery
B.
Open the cover by first unscrewing the two screws on the far right of the
front cover.
C.
Locate the four position terminal strip with jumper(s). Note: There is a red
stripe on top of each jumper in the photo below.
D.
Two jumpers are used for 115VAC input and one jumper (across the middle
two screws) is used for 230VAC input. See Figures below.
Jumpers
115 volt AC setting
230 volt AC setting
E.
For 230 VAC operation, the 115 VAC cord must be replaced with one for
230 VAC in required country configuration.
NOTE: If the plug has to be changed make sure to connect
the green AC line wire to ground.