SETF
Output Field Setting Command
Input:
SETF <field> [term]
Format:
±
nnn.nnnE
±
nn
<field>
Specifies the output field setting: 0.0000 –
±
601.000E+03 G or 0.0000 –
±
60.1000E+00 T.
Remarks:
Sets the field value that the output will ramp to at the present ramp rate. The setting entered will be
based on the field constant and the field units. Refer to the FLDS command.
SETF?
Output Field Setting Query
Input:
SETF? [term]
Returned:
<field> [term]
Format:
±
nnn.nnnE
±
nn (Refer to command for description)
SETI
Output Current Setting Command
Input:
SETI <current> [term]
Format:
±
nn.nnnn
<current>
Specifies the output current setting: 0.0000 –
±
60.1000A.
Remarks:
Sets the current value that the output will ramp to at the present ramp rate.
Setting value is limited by LIMIT.
SETI?
Output Current Setting Query
Input:
SETI? [term]
Returned:
<current> [term]
Format:
±
nn.nnnn
(Refer to command for description)
SETV
Output Compliance Voltage Setting Command
Input:
SETV <voltage> [term]
Format:
+
n.nnnn
<voltage>
Specifies the output compliance voltage setting: 0.1000 – 5.0000 V.
Remarks:
Sets the output compliance voltage. This value will be used in both the positive and negative
directions. Setting value is limited by LIMIT.
SETV?
Output Compliance Voltage Setting Query
Input:
SETV? [term]
Returned:
<voltage> [term]
Format:
+
n.nnnn (Refer to command for description)
STOP
Stop Output Current Command
Input:
STOP [term]
Remarks:
This command will stop the output current ramp within two seconds of sending the command. To
restart the ramp, use the
SETI
command to set a new output current setpoint or the SETF command to
set a new output field setpoint.
Содержание 625
Страница 2: ...Lake Shore Model 625 Superconducting MPS User s Manual A ...
Страница 5: ...Lake Shore Model 625 Superconducting MPS User s Manual D ...
Страница 6: ...Lake Shore Model 625 Superconducting MPS User s Manual C ...
Страница 8: ...Lake Shore Model 625 Superconducting MPS User s Manual E ...
Страница 20: ...Lake Shore Model 625 Superconducting MPS User s Manual 1 8 Introduction This Page Intentionally Left Blank ...
Страница 36: ...Lake Shore Model 625 Superconducting MPS User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 100: ...Lake Shore Model 625 Superconducting MPS User s Manual 6 2 Options and Accessories ...
Страница 101: ...Lake Shore Model 625 Superconducting MPS User s Manual Options and Accessories 6 3 This Page Intentionally Left Blank ...
Страница 102: ......