Lake Shore Model 625 Superconducting MPS User’s Manual
Installation
3-1
CHAPTER 3
INSTALLATION
3.0
GENERAL
This chapter provides general installation instructions for the Model 625 Superconducting Magnet Power Supply. Please
read this entire chapter before installing the instrument and powering it on to ensure the best possible performance and
maintain operator safety. For instrument operating instructions refer to Chapter 4. For computer interface installation and
operation refer to Chapter 6.
NOTE:
It is recommended that the instrument be powered up and operated with the shorting bar in place before
connecting it to a magnet. This will allow the user to setup the supply and become comfortable with its
operation without risk of damage to the magnet.
3.1
INSPECTION AND UNPACKING
Inspect shipping containers for external damage before opening them. Photograph any container that has significant
damage before opening it. If there is visible damage to the contents of the container contact the shipping company and
Lake Shore immediately, preferably within 5 days of receipt of goods. Keep all damaged shipping materials and contents
until instructed to either return or discard them.
Open the shipping container and keep the container and shipping materials until all contents have been accounted for.
Check off each item on the packing list as it is unpacked. Instruments themselves may be shipped as several parts. The
items included with the Model 625 are listed below. Contact Lake Shore immediately if there is a shortage of parts or
accessories. Lake Shore is not responsible for any missing items if not notified within 60 days of shipment.
Inspect all items for both visible and hidden damage that occurred during shipment. If damage is found, contact Lake
Shore immediately for instructions on how to file a proper insurance claim. Lake Shore products are insured against
damage during shipment but a timely claim must be filed before Lake Shore will take further action. Procedures vary
slightly with shipping companies. Keep all shipping materials and damaged contents until instructed to either return or
discard them.
If the instrument must be returned for recalibration, replacement or repair, a returned goods (RA) number must be
obtained from a factory representative before it is returned. The Lake Shore RA procedure is given in Paragraph 7.2.
Items Included with Model 625 Superconducting Magnet Power Supply
1
Model 625 Instrument
1
Model 625 User’s Manual
2
Front Handles
2
Rear Handles
1
Digital I/O Mating Connector
1
Analog I/O Mating Connector
1
Output Shorting Bar and Terminal Fasteners (Installed on output bus bars)
1
Line Power Cord
1
Fuse Pair for Alternative Voltage *
1
Line Power Cord for Alternative Voltage *
*
Included only when purchased with VAC-120-ALL Power Option.
3.1.1
Moving and Handling
Four handles are provided for ease of moving and handling the Model 625. Always use at least two, and if possible four,
handles when carrying the unit.
Содержание 625
Страница 2: ...Lake Shore Model 625 Superconducting MPS User s Manual A ...
Страница 5: ...Lake Shore Model 625 Superconducting MPS User s Manual D ...
Страница 6: ...Lake Shore Model 625 Superconducting MPS User s Manual C ...
Страница 8: ...Lake Shore Model 625 Superconducting MPS User s Manual E ...
Страница 20: ...Lake Shore Model 625 Superconducting MPS User s Manual 1 8 Introduction This Page Intentionally Left Blank ...
Страница 36: ...Lake Shore Model 625 Superconducting MPS User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 100: ...Lake Shore Model 625 Superconducting MPS User s Manual 6 2 Options and Accessories ...
Страница 101: ...Lake Shore Model 625 Superconducting MPS User s Manual Options and Accessories 6 3 This Page Intentionally Left Blank ...
Страница 102: ......