Lake Shore Model 625 Superconducting MPS User’s Manual
Computer Interface Operation
5-41
QNCH?
Quench Parameter Query
Input:
QNCH? [term]
Returned:
<enable>, <rate> [term]
Format:
n,
+
n.nnnn
(Refer to command for description)
RATE
Output Current Ramp Rate Setting Command
Input:
RATE <rate> [term]
Format:
+n.nnnn
<rate>
Specifies the rate at which the current will ramp at when a new output current setting is
entered: 0.0001 – 99.999 A/s.
Remarks:
Sets the output current ramp rate. This value will be used in both the positive and negative directions.
Setting value is limited by LIMIT.
RATE?
Output Current Ramp Rate Setting Query
Input:
RATE? [term]
Returned:
<rate> [term]
Format:
+n.nnnn (Refer to command for description)
RATEP
Persistent Mode Ramp Rate Parameter Command
Input:
RATEP <enable>, <rate> [term]
Format:
n, +n.nnnn [term]
<enable>
Specifies if the persistent mode ramp rate is to be used when the magnet is in persistent
mode (PSH heater off): 0 = Never use persistent mode ramp rate, 1 = Use persistent mode
ramp rate when in persistent mode.
<rate>
Specifies the ramp rate to use while in persistent mode: 0.0001 – 99.999 A/s.
Remarks:
Typically the current can be ramped faster when the magnet is in persistent mode since the current
change is not seen by the inductance of the magnet. This setting will automatically change the ramp
rate when the magnet goes into or out of persistent mode.
RATEP?
Persistent Mode Ramp Rate Parameter Query
Input:
RATEP? [term]
Returned:
<enable>, <rate> [term]
Format:
n,
+
n.nnnn
(Refer to command for description)
RDGF?
Field Output Reading Query
Input:
RDGF? [term]
Returned:
<field> [term]
Format:
±
nn.nnnnE
±
nn
<field>
Calculated output field reading. Field is calculated by multiplying the field constant by
the measured current output. Use the
FLDS
command to set the field constant and field
constant units (G or T).
Содержание 625
Страница 2: ...Lake Shore Model 625 Superconducting MPS User s Manual A ...
Страница 5: ...Lake Shore Model 625 Superconducting MPS User s Manual D ...
Страница 6: ...Lake Shore Model 625 Superconducting MPS User s Manual C ...
Страница 8: ...Lake Shore Model 625 Superconducting MPS User s Manual E ...
Страница 20: ...Lake Shore Model 625 Superconducting MPS User s Manual 1 8 Introduction This Page Intentionally Left Blank ...
Страница 36: ...Lake Shore Model 625 Superconducting MPS User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 100: ...Lake Shore Model 625 Superconducting MPS User s Manual 6 2 Options and Accessories ...
Страница 101: ...Lake Shore Model 625 Superconducting MPS User s Manual Options and Accessories 6 3 This Page Intentionally Left Blank ...
Страница 102: ......