Lake Shore Model 625 Superconducting MPS User’s Manual
7-20
Service
Calibration-Specific Interface Commands (Continued)
CALG?
Gain Calibration Constant Query
Input:
CALG? <type>[term]
Format:
nn
<type>
1 – 10.
Returned:
<value>[term]
Format:
±nnnnnnn
(Refer to command for description)
CALSAVE
Calibration Save Command
Input:
CALSAVE[term]
Remarks:
Saves all CALZ and CALG calibration constants in non-volatile memory.
CALZ
Zero Offset Calibration Constant Command
Input:
CALZ <type>, <value>[term]
Format:
nn, ±nnnnnnn
<type>
Specifies the item to calibrate. Valid entries are:
0 = Output Current Reading
1 = Bias A Reading *
2 = Bias B Reading *
3 = CM Voltage Reading *
4 = Out Control Reading *
5 = Output Voltage Reading
6 = Remote Voltage Reading
7 = External Programming Voltage
8 = Temperature *
9 = Raw Supply Voltage *
10 = Actual Output Current
<value>
Zero offset calibration constant value.
Remarks:
Items marked with a * are for internal diagnostic use only and should always be set to a value of 0
(default).
CALZ?
Zero Offset Calibration Constant Query
Input:
CALZ? <type>[term]
Format:
nn
<type>
1 – 10.
Returned:
<value>[term]
Format:
±nnnnnnn
(Refer to command for description)
Содержание 625
Страница 2: ...Lake Shore Model 625 Superconducting MPS User s Manual A ...
Страница 5: ...Lake Shore Model 625 Superconducting MPS User s Manual D ...
Страница 6: ...Lake Shore Model 625 Superconducting MPS User s Manual C ...
Страница 8: ...Lake Shore Model 625 Superconducting MPS User s Manual E ...
Страница 20: ...Lake Shore Model 625 Superconducting MPS User s Manual 1 8 Introduction This Page Intentionally Left Blank ...
Страница 36: ...Lake Shore Model 625 Superconducting MPS User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 100: ...Lake Shore Model 625 Superconducting MPS User s Manual 6 2 Options and Accessories ...
Страница 101: ...Lake Shore Model 625 Superconducting MPS User s Manual Options and Accessories 6 3 This Page Intentionally Left Blank ...
Страница 102: ......