64
c
HAPTER
4:
Operation
Model 350 Temperature Controller
D
Input and Units:
each available display location has an associated Input and Units
setting. The Input parameter determines which sensor will be used as the input
of the displayed data. The Input can be any of the four sensor inputs, or None. If
None is selected, then the display location will be blank. The Units parameter
determines which units to display the reading in.
Menu Navigation:
Display Setup
Q
Location
(1, 2, 3, 4, 5, 6, 7, 8)
Input
Q
(None, Input A, Input B,
Input C, Input D)
Display Setup
Q
Location
(1, 2, 3, 4, 5, 6, 7, 8)
Units
Q
(Kelvin, Celsius, Sensor,
Min, Max, Sensor Name)
Interface Command:
DISPFLD
Defaults
Location
Input
Units
1
Input A
Kelvin
2
Input B
3
Input C
4
Input D
5
Input A
Sensor
6
Input B
7
Input C
8
Input D
D
Displayed Output:
in the Custom Display mode the bottom half of the display is
dedicated for output and control loop information for one of the four outputs.
The source of this information depends on the output selected for the Displayed
Output parameter. If the selected output is configured as a control loop output,
then all associated control loop parameters will be displayed.
When viewing the Custom Display screen, the configured Displayed Output is
signified by L1, L2, L3 or L4, followed by the control loop input, if applicable. The L
character stands for Loop, but will be displayed even for outputs that are not con-
figured as control loop outputs.
Menu Navigation:
Display Setup
Q
Displayed Output
Q
Output (1, 2, 3, 4)
D
Default: Output 1
Interface Command:
DISPLAY
TABLE 4-6
Содержание 350
Страница 4: ...Model 350 Temperature Controller ...
Страница 6: ......
Страница 14: ...Model 350 Temperature Controller ...
Страница 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Страница 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Страница 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Страница 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Страница 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Страница 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...