3.4.3 Power Cord
47
|
www.lakeshore.com
3.4.3 Power Cord
The Model 350 includes a 3-conductor power cord that mates with the IEC 320-C14
line cord receptacle. Line voltage is present on the two outside conductors and the
center conductor is a safety ground. The safety ground attaches to the instrument
chassis and protects the user in case of a component failure. A CE approved power
cord is included with instruments shipped to Europe; a domestic power cord is
included with all other instruments (unless otherwise specified when ordered).
Always plug the power cord into a properly grounded receptacle to ensure safe instru-
ment operation. Do not use a detachable mains supply cord with inadequate rating.
Position the Model 350 in such a way to enable easy access to the disconnecting device.
Failure to comply could result in death or injury to personnel.
If the power supply cord is damaged or lost, it must be replaced. Contact Lake Shore for a
replacement to ensure proper voltage, current and type of cord. The power supply cord
must not exceed 3 m in length.
The delicate nature of measurements being taken with this instrument may necessi-
tate additional grounding including ground strapping of the instrument chassis. In
these cases the operators safety should remain the highest priority and low imped-
ance from the instrument chassis to safety ground should always be maintained.
3.4.4 Power Switch
The power switch is part of the line input assembly on the rear panel of the Model 350
and turns line power to the instrument on and off. When the circle is depressed,
power is off. When the line is depressed, power is on.
3.5 Diode/Resistor
Sensor Inputs
This section details how to connect diode and resistive sensors to the Model 350
inputs. Refer to section 4.4 to configure the inputs. Refer to section 3.5.8 through sec-
tion 3.5.9 for a description of the optional capacitance input and thermocouple input.
3.5.1 Sensor Input
Connector and Pinout
The input connectors are 6-pin DIN 45322 sockets. The sensor connector pins are
defined in FIGURE 3-3 and TABLE 3-2. Four mating connectors (6-pin DIN plugs) are
included in the connector kit shipped with the instrument. These are common con-
nectors, so additional mating connectors can be purchased from local electronics
suppliers. They can also be ordered from Lake Shore as G-106-233.
Sensor input connector
FIGURE 3-3
Содержание 350
Страница 4: ...Model 350 Temperature Controller ...
Страница 6: ......
Страница 14: ...Model 350 Temperature Controller ...
Страница 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Страница 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Страница 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Страница 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Страница 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Страница 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...