152
c
HAPTER
6:
Computer Interface Operation
Model 350 Temperature Controller
OPSTE
Operational Status Enable Command
Input
OPSTE <bit weighting> [term]
Format
nnn
Remarks
Each bit has a bit weighting and represents the enable/disable mask of the corre-
sponding operational status bit in the Operational Status Register. This determines
which status bits can set the corresponding summary bit in the Status Byte Register.
To enable a status bit, send the command OPSTE with the sum of the bit weighting for
each desired bit. Refer to section 6.2.5.2 for a list of operational status bits.
OPSTE?
Operational Status Enable Query
Input
OPSTE?[term]
Returned
<bit weighting> [term]
Format
nnn (Refer to section 6.2.5.2 for a list of operational status bits)
OPSTR?
Operational Status Register Query
Input
OPSTR? [term]
Returned
<bit weighting> [term]
Format
nnn
Remarks
The integers returned represent the sum of the bit weighting of the operational sta-
tus bits. These status bits are latched when the condition is detected. This register is
cleared when it is read. Refer to section 6.2.5.2 for a list of operational status bits.
OUTMODE
Output Mode Command
Input
OUTMODE <output>,<mode>,<input>,<powerup enable>[term]
Format
n,n,n,n
<output>
Specifies which output to configure: 1–4.
<mode>
Specifies the control mode. Valid entries: 0 = Off, 1 = Closed
Loop PID, 2 = Zone, 3 = Open Loop, 4 = Monitor out,
5 = Warmup Supply
<input>
Specifies which input to use for control: 0 = None, 1 = A,
2 = B, 3 = C, 4 = D (5 = Input D2, 6 = Input D3, 7 = Input D4,
8 = Input D5 for 3062 option)
<powerup enable>
Specifies whether the output remains on or shuts off after
power cycle. Valid entries: 0 = powerup enable off,
1 = powerup enable on.
Example
OUTMODE 1,2,1,0[term]
—Output 1 configured for Zone control mode, using Input A
for the control input sensor, and will turn the output off when power is cycled.
Remarks
Modes 4 and 5 are only valid for Analog Outputs (3 and 4).
OUTMODE?
Output Mode Query
Input
OUTMODE? <output>[term]
Format
n
<output>
Specifies which output to query: 1–4.
Returned
<mode>,<input>,<powerup enable>[term]
Format
n,n,n (refer to command for description)
Содержание 350
Страница 4: ...Model 350 Temperature Controller ...
Страница 6: ......
Страница 14: ...Model 350 Temperature Controller ...
Страница 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Страница 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Страница 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Страница 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Страница 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Страница 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...