percentages. Note that selecting the correct coolant is the responsibility of the customer and should
match what is filled in the system by the customer. Selecting the wrong coolant may cause damage to
the equipment.
Water: 10
°
C to 40
°
C
10% Ethylene Glycol: 5
°
C to 40
°
C
20% Ethylene Glycol: 0
°
C to 40
°
C
30% Ethylene Glycol: -5
°
C to 40
°
C
Figure 26: Choosing the Coolant
13.6.3 Boost mode
Figure 27 Boost mode
The unit comes with an option to increase chiller performance using Boost Mode, if required. With the
Boost Mode ON, the fans run at a higher speed and increase the cooling capacity of the unit. Boost
Mode is turned OFF by default. To activate Boost Mode, press ON and store the selection by pressing
the green arrow button.
13.6.4 Customer Configurable Alarms
Figure 28: Setpoint Alarms