19
Com. Channel
To be able to operate several VitalSim units in the same vicinity,
it is possible to select 5 different communication channels.
Factory setting is channel 1.
Communication channels can only be changed if the remote
control and base unit are connected via cable.
Calibrate BP
To ensure correct measurements of simulated BP,
the sphygmomanometer used with the cuff needs to be
calibrated to the pressure sensor in the VitalSim.
This is done by selecting the
Calibrate BP
function, inflating
the cuff to 100 mmHg and pressing the
Calibrate
quick
execution button as the pressure is held at exactly 100 mmHg.
Language
One of several languages can be selected. Language selection
affects the texts on the remote control display. Selection of a
new language will also affect the
“Yes”
and
“No”
vocal sounds.
Factory setting is English