Performing Simulations
Perf
orming
Sim
ulations
Catheter Installation
SimMom includes two types of urinary connectors. The first type is
for scenarios involving intermittent catheterization. This is referred
to as the intermittent catheter tube. The second is a bag-type
connector for scenarios involving an indwelling balloon catheter. This
is referred to as the indwelling catheterization bladder. Except for the
extra attachment flap on the bag, both connectors are removed and
attached in the same way:
1. Disconnect the white Luer connector from the urine reservoir
outlet. Ensure reservoir is empty or the tube is clamped.
2. Pull the tube free of locating groove on pelvic ring clamp.
3. Remove the pelvic ring clamp.
4. Pull the urinary connector from spigot on grey urinary valve, using
a sideways “rocking” action. Avoid pulling directly because the
urinary valve can come away from the urethral tube.
Note:
If the urinary valve detaches from the urethral tube,
simply push it back on.
5. Grip the grey urinary value through the birth canal skin, and push a
replacement urinary connector onto valve spigot. Ensure that the
connector tube is pointing to the manikin’s right side.
6. Push the connector tube into the groove toward the pelvic ring.
7. Connect the white Luer to the urine reservoir outlet tube.
8. For cord prolapse scenarios, use the indwelling catheterization
bladder. Push hole on the flap of the bladder over the retaining
lug on the inside of the abdominal skin.
– For PPH scenarios, fold flap back and tuck behind indwelling
catheterization bladder.
9. Replace pelvic ring clamp.
To remove and replace the urinary valve:
1. Remove pelvic ring clamp.
2. Remove urinary connector.
3. Pull grey urinary valve from urethral tube on birth canal.
4. Push outlet spigot (wider) of replacement valve into urethral tube.
Grip tube through birth canal skin to provide support.
5. Push urinary connector onto inlet spigot (narrower).
6. Replace pelvic ring clamp.
Filling Blood and Urine Reservoirs
Reservoirs can be filled while disconnected from SimMom or while
in position. The urine reservoir contains a maximum of 400 mL. The
blood reservoir contains a maximum of 800 mL. Both the urine and
blood reservoirs are filled in the same way.
1. Remove red filler cap.
2. If filling outside the model, ensure the slide clamp is in the closed
position.
3. Stabilize the reservoir by firmly gripping the filling port.
4. Open up the bag by pushing a finger in through the filler hole and
separating the two walls.
Содержание SimMom
Страница 1: ...SimMom Directions for Use www laerdal com www limbsandthings com ...
Страница 2: ......
Страница 43: ...Introduction Setup Software Maintenance Spare Parts Specifications 43 Maintenance Performing Simulations ...
Страница 44: ...44 44 Notes ...
Страница 46: ...46 46 Notes ...
Страница 48: ...48 Notes 48 ...
Страница 49: ......
Страница 50: ... 2012 Laerdal Medical All rights reserved 20 05176 Rev B Printed in USA ...