13
MOTHLY MAINTENANCE
Regular Cleaning of Fluid and Blood System
Regular cleaning of the Fluid and Blood System is recommended as part
of proper care of the product. Once or twice a month, the Fluid and
Blood System should be cleaned thoroughly.
Cleaning Fluid System
1.
Ensure that the simulator’s power is ON.
2.
Ensure that the simulator’s internal fluid reservoir is drained. See:
Empty Internal Fluid Reservoir.
3.
Press the fill button located on the fill panel. The LED indicator on
the fill button will light up.
4.
Connect an empty fill bottle to the fluid and air connector in the fill
panel, and the filling of air into the internal reservoir will start.
5.
We recommend covering the simulator’s head with a towel, as
pumping air through the system might cause some squirting.
6.
With the fill button activated, open the Circulation and Fluids tab in
LLEAP and check the boxes for Sweat, Ears, Eyes, Nose, Mouth and
Urine (polyuria).
7.
Wait until there are no more fluids emitted from the simulator and
then uncheck all boxes.
8.
Disconnect the empty fill bottle.
Flush the system with isopropanol
9.
Connect a fill bottle with isopropanol to the fluid and air connector
in the fill panel, and the filling of isopropanol into the internal
reservoir will start.
10.
In LLEAP, again check the boxes for Sweat, Ears, Eyes, Nose, Mouth
and Urine (polyuria).
11.
Wait until isopropanol is emitted from all the fluid outlets.
Содержание SimMan 3G
Страница 1: ...Laerdal SimMan 3G Maintenance Guide Tasmanian Simulation Collaborative February 2016...
Страница 4: ...4 OVERVIEW RFID Antennae Bleeding ports Speakers ECG Connectors Defib Connectors Pulses...
Страница 5: ...5 SETUP WIRELESS Headset Instructor PC SpO2 Patient Monitor USB cable...
Страница 6: ...6 SETUP WIRED Headset Instructor PC SpO2 Patient Monitor USB cable Ethernet cable Switch...
Страница 16: ...16 www tcen com au tcen dhhs tas gov au 03 6233 2756...