MAINTENANCE

              

   

MAINTENANCE

              

Replacing Crico Tape/Neck Skin

After creating an emergency airway through the cricothyroid 
membrane, replace the perforated membrane before starting a new 
simulation session.

1  Remove the neck skin (hook and loop fasteners  

behind the neck).

2  Remove the old strip of Crico tape.

3  Replace with a new Crico tape.

4  Ensure Crico Tape completely covers and seals the opening to 

prevent leakage while ventilating the Patient Simulator.

Replacing Chest Drain Pleura

The chest drain module’s pleura skin should be replaced after  
each use. 

1  Open the Torso Skin and remove the module from the chest.

2  Remove the old pleura skin, and replace with a new skin and 

replace the module.

Replacing Pneumothorax Bladders

After multiple pneumothorax decompressions, 

the bladders may need replacement:

1  Open the Torso Skin to expose the chest 

plate. Lift the chest plate to reveal the 
pneumothorax bladders located in slots in the 
side of the chest plate assembly.

2  Slide out the used pneumothorax bladder.

3  Disconnect the tube and discard the old 

bladder.

4  Insert the new bladder into the slot.

5  Reconnect the tube to the new bladder.

Replacing Chest Rise Bladder

 If the chest rise bladders leak or are damaged:

1  Open the Torso Skin to expose the chest plate.  

There is one bladder on each side of the chest 
plate assembly. 

2  Disconnect the tube from the bladder.

3  Discard the used bladder.

4  Insert new bladder.

5  Reconnect the tube to the new bladder.

Left Leg to Pelvis – Tube Descriptions

Name/Label

Tube Color

Connector Description

Pneum L

Silicon

Barb connector

Pneum R

Silicon

Barb connector

Chest L

Silicon

Barb connector

Chest R

Silicon

Barb connector

Cautions 

and 

W

arnings

Featur

es

Setup

Maintenance

Spar

e Par

ts

Tr

oubleshooting

Содержание SimMan 3G

Страница 1: ...SimMan 3G User Guide www laerdal com EN...

Страница 2: ...LLEAP 22 RFIDTag Setup 23 Using RFIDTags 23 Connecting Defibrillation Adapter Plates 24 Connecting the Blood Pressure Cuff 25 Calibrating the Blood Pressure Cuff Using LLEAP 25 Connecting the SpO2 Pro...

Страница 3: ...the internal tubing and cabling is disconnected Never use the Patient Simulator outdoors in wet conditions as this may pose a shock hazard or damage the simulator Never use the Patient Simulator in te...

Страница 4: ...ulation session comprises the following main applications LLEAP Laerdal Learning Application Voice Conference Application Patient Monitor SimView Server or SessionViewer SimDesigner and other applicat...

Страница 5: ...geal swelling Laryngospasm Decreased cervical range of motion Trismus Teeth soft upper dentures may be replaced with a hard set of teeth for enhanced realism while practicing intubations During simula...

Страница 6: ...can be made See LLEAP Help for further information Sounds Two types of sounds can be used in a scenario Body sounds Vocal Sounds The sounds can be triggered by the scenario or controlled by the instr...

Страница 7: ...the clips on either side 2 Insert both batteries into the battery tray 3 Snap the battery clamp back into place over the batteries 4 Connect the corresponding battery cables from the batteries to the...

Страница 8: ...arging time is approximately 3 hours The external battery charger should only be used with SimMan 3G batteries Charger light showing battery sign Light Code Light Color Characteristic Standby Yellow S...

Страница 9: ...Patient Simulator is stationary over extended periods of use If the Patient Simulator is required to exhale CO2 with each ventilation connect external CO2 Connect CO2 only if the Patient Simulator is...

Страница 10: ...he button will light and blood will flow into the Patient Simulator 4 Charge the system for 60 seconds before starting bleeding simulation Warning Connecting a full blood fill unit to a Patient Simula...

Страница 11: ...this process for as many times as is necessary Note If the external fill unit runs empty during a bleeding scenarios air will be introduced to the blood system causing inaccurate reading The IV Cathe...

Страница 12: ...carry an RFID tag RFID tagging Syringes Airway Devices The RFID tags come as part of the SimMan 3G Drug and Airway Tag Kit Tags are pre printed with information to identify drugs drug concentration or...

Страница 13: ...not defibrillate the Patient Simulator without the torso skin Warning Do not defibrillate the Patient Simulator in a flammable or oxygen enriched atmosphere Warning The Patient Simulator torso must a...

Страница 14: ...e soft set may be replaced with a hard set of teeth 1 Remove the teeth from the mouth 2 Align the new set of teeth with the gums and push them back until the teeth engage and lock onto the gums 3 Ensu...

Страница 15: ...default leg OpenTorso Skin and stomach foam as shown in Maintenance 2 Unscrew hip joint connector Remove the leg cables and tubes from the connector 3 Carefully remove the left leg with cables and tub...

Страница 16: ...obility Note The SimMan 3G System exceeds the weight allowance on most commercial airlines Some parts may have to be transported separately For more information on weight restrictions contact the rele...

Страница 17: ...flushed with isopropanol press fill the button again to deactivate the filling of isopropanol to the reservoir The LED indicator on the fill button will now be off 13 Leave the isopropanol fill bottl...

Страница 18: ...a new router attach the power cable to the DC plug and the Ethernet cables back into the router again Opening theTorso Open the Patient Simulator torso for the following procedures Attaching or Repla...

Страница 19: ...5 Tighten the shoulder screw with the Allen wrench LAPULSES BP LACONV P B L BP PULSES L L LA B 6 Connect the arm cables to the corresponding connection points in the torso Left Arm toTorso Cable andT...

Страница 20: ...ns the bladders may need replacement 1 Open theTorso Skin to expose the chest plate Lift the chest plate to reveal the pneumothorax bladders located in slots in the side of the chest plate assembly 2...

Страница 21: ...skins easier To prevent the zippers from separating from the skin make sure to properly position the skin halves and hold them together while zipping them closed Replacing Blood System Filter If reduc...

Страница 22: ...d or are leaking See Opening the Torso In case of fluid leakage power off the Patient Simulator and contact LaerdalTechnical Service Problem Identifying a single Patient Simulator when there are multi...

Страница 23: ...nt Simulator internal pkg 2 212 12350 Thigh joint connector 212 12450 Battery clip Patient Simulator internal 212 12550 Panel cover left and right 381107 SoftTeeth 212 14250 Arm Adaptor SimMan SimMan...

Страница 24: ...20 07890 Rev D 2016 Laerdal Medical AS All rights reserved Manufacturer Laerdal Medical AS P O Box 377 Tanke Svilandsgate 30 4002 Stavanger Norway T 47 51 51 17 00 www laerdal com...

Отзывы: