Maintaining Arm IO Bone
1. Unzip arm skin from shoulder to wrist.
2. Remove Arm IM Pad.
3. Lift and slide out the used Arm IO Bone.
4. Puncture holes in the used IO bone can be repaired with
Loctite 454 or similar superglue
OR
replace the IO bone
completely when it has been significantly punctured.
5. Replace repaired or new Arm IO Bone.
6. Ensure that base of IO Bone slots into place.
7. Replace Arm IM Pad.
Caution
If using fluids, ensure that the bone has been repaired properly, or
use a new bone, to avoid causing internal leakage and damage.