20-18071
Rev A
Charging the Simulator
Battery
SimMan 3G PLUS is powered by 2 Laerdal Lithium Ion (Li-Ion)
batteries. Always use 2 batteries together to power the
Simulator and ensure that they are connected properly.
Refer to
Setup Batteries
.
Charging
• Battery charging time is approximately 3 hours. Once fully
charged, the batteries can last for up to 4 hours.
• Connect the Patient Simulator to the external power
supply. Turn the Patient Simulator on to charge the
batteries.
• Battery and power status can be viewed on the Patient
Simulator’s power panel and the Simulator Status window
in LLEAP.
Notes
•
On approximately every 30th charge cycle, drain the battery
completely before recharging. To drain the batteries, run the
Patient Simulator on both batteries until automatic shutdown.
•
The Patient Simulator will automatically shut down if battery
temperature rises above 60 °C (140 °F) or the remaining
charge falls below 6% on one of the two batteries.
•
If both batteries are removed while the simulation is paused,
the Patient Simulator will shut down and simulation data will
be lost.
Cautions
•
Do not run the Patient Simulator for more than 1 minute on a
single battery.
•
After the Patient Simulator is turned off, wait 20 seconds
before restarting or the Patient Simulator may not function
properly.
•
Never store fully charged batteries for longer than a month.
•
Only use the SimMan Family (except SimMan ALS) external
power supply and batteries.
•
The external battery charger is for indoor use only.
•
Never store the batteries inside the Patient Simulator.
Refer to the
Important Product Information
booklet for
further information on battery care and specifications.
Содержание SimMan 3G PLUS
Страница 1: ...20 18071 Rev A SimMan 3G PLUS User Guide...
Страница 2: ...20 18071 Rev A SimMan 3G PLUS Introduction...
Страница 7: ...20 18071 Rev A SimMan 3G PLUS Features Overviews...
Страница 10: ...20 18071 Rev A Panels Overview 1 Power Panel 2 Fluid Fill Panel 3 Air CO2 Panel...
Страница 12: ...20 18071 Rev A Fluid Panel Overview 1 Air Out 2 Blood Inlet 3 Fluid Inlet 4 Activate Filling...
Страница 22: ...20 18071 Rev A SimMan 3G PLUS Pre Simulation Setup General Pre Simulation Setup PrepareTrauma Leg Accessory...
Страница 39: ...20 18071 Rev A SimMan 3G PLUS Immersive Simulation...
Страница 41: ...20 18071 Rev A SimMan 3G PLUS Intravenous IV Injection...
Страница 45: ...20 18071 Rev A SimMan 3G PLUS Intraosseous IO Injection...
Страница 52: ...20 18071 Rev A SimMan 3G PLUS Breathing Interventions...
Страница 55: ...20 18071 Rev A SimMan 3G PLUS Cardiac Interventions...