SimMan 3G PLUS enables bilateral, automatic NIBP monitoring
with a clinical monitor.
Note
Only use the Blood Pressure cuff supplied with SimMan 3G PLUS.
1. Locate correct connector (supplied) to connect BP cuff to
Patient Monitor.
2. Attach connector to BP cuff (use a Y-connector if the
patient monitor uses a dual tube).
3. Connect short tube with angled connector to BP port
and longer tube to the Patient Monitor.
The NIBP system delivers simulated blood pressures with
an accuracy of ≤ 5 mmHg or 3% of the reading whichever is
greater (at 15-25 °C, 15-85% RH).
NIBP system range
Diastolic blood pressure
40-130 mmHg
Systolic blood pressure
60-200 mmHg
Refer to
LLEAP Help Files
for complete blood pressure
measurements and operating information.
Содержание SimMan 3G PLUS
Страница 1: ...20 18071 Rev A SimMan 3G PLUS User Guide...
Страница 2: ...20 18071 Rev A SimMan 3G PLUS Introduction...
Страница 7: ...20 18071 Rev A SimMan 3G PLUS Features Overviews...
Страница 10: ...20 18071 Rev A Panels Overview 1 Power Panel 2 Fluid Fill Panel 3 Air CO2 Panel...
Страница 12: ...20 18071 Rev A Fluid Panel Overview 1 Air Out 2 Blood Inlet 3 Fluid Inlet 4 Activate Filling...
Страница 22: ...20 18071 Rev A SimMan 3G PLUS Pre Simulation Setup General Pre Simulation Setup PrepareTrauma Leg Accessory...
Страница 39: ...20 18071 Rev A SimMan 3G PLUS Immersive Simulation...
Страница 41: ...20 18071 Rev A SimMan 3G PLUS Intravenous IV Injection...
Страница 45: ...20 18071 Rev A SimMan 3G PLUS Intraosseous IO Injection...
Страница 52: ...20 18071 Rev A SimMan 3G PLUS Breathing Interventions...
Страница 55: ...20 18071 Rev A SimMan 3G PLUS Cardiac Interventions...