(B) Mid-Axillary Site (RIGHT):

1 .  Remove chest skin from tabs at the shoulder and back .
2 .  Remove pneumo bladder box from right side of torso . (Figure 4) 

3 .  Disconnect bladder hose from inline hose connector, being sure hose does 

not fall through hole and into torso .

4 .  Remove bladder from pneumo pad and discard .

5.  Trim bladder tubing to match original tubing length to fit this manikin style. 

Connect new bladder hose to inline hose connector .

6 .  Fold and insert new pneumo bladder into pneumo box .
7 .  Replace chest skin over torso, being sure to secure skin at shoulder and back 

tabs .

8 .  Fill puncture marks on exterior of chest skin with the Bladder Replacement 

Kit .

NOTE:

 Bladders can be repaired using vulcanized or contact cement glue (not 

provided) . Cover punctured surface of pneumothorax bladders with glue . Allow 

bladders to air dry completely overnight before replacing in the chest plate .

Chest Drain:

Mid Axillary Site (LEFT)

1 .  Remove chest skin from tabs at shoulders and back .
2 .  Remove the chest tube drain module from the left mid-axillary site of the 

manikin . (Photo 14) 

3 .  Insert a new chest drain module .
4 .  Replace the chest skin over the torso being sure to secure the skin at the 

shoulders and both sides .

Maintenance:

Cleaning

1 .  Clean with mild soap and water; do not submerse the manikin or parts in 

cleaning fluids or water.

2 .  Use only on clean surface . Avoid felt tipped markers, ink pens, acetone, iodine 

or other staining products and avoid placing the manikin on newsprint or 
inked lines of any kind .

3 .  To ensure longevity, each manikin should be cleaned after each training 

session and a general inspection should be conducted regularly .

4 .  Modules and all other parts should be drained and air-dried thoroughly 

before storage and disinfected when needed . After use of injection pads (use 
water only), accumulated water should be squeezed out . 

Do not store wet 

foam pads in the skin.

 To prevent mildew or mold, pads can be soaked in 

a mild solution of disinfectant and water or bleach and water . Squeeze excess 
solution from pads, allow them to dry, then store or reinsert in manikin .

5.  Articulating parts will benefit from a light application of talcum powder prior 

to training sessions .

To extend the cleanliness of your manikin:

1 .  Use a small towel and lightly dust the manikin head and chest with a small 

amount of powder (provided) . Wipe away any excess .

2 .  Use gloves when handling the manikin .
3 .  Avoid felt tipped markers, ink pens, iodine and newsprint, these will stain the 

manikin .

Laerdal Recommends

Use only on clean surface . Avoid felt tipped markers, ink pens, acetone, iodine 

or other staining products and avoid placing the manikin on newsprint or 

inked lines of any kind . These will stain the manikin .

Attaching/Removing arms:

1 .  Remove deltoid injection pad from upper arm .
2 .  Detach chest skin from tabs at shoulders and back .
3 .  Remove skin to reveal internal portion of upper chest .
4 .  Lift hard chest plate to reveal inside sockets for arm connections .

5.  Place one washer on threaded 4 ½” bolt.

6 .  Insert threaded bolt through upper arm and through torso so that it is visible 

in chest cavity . (Figure 5) 

7 .  From inside of chest area, slip a washer, a spring and another washer on bolt .
8 .  Screw a wing nut on bolt and tighten until desired articulation is achieved .
9 .  Replace hard chest plate .

10.  Pull chest skin back onto shoulder area.

11 .  Connect chest skin to torso at shoulder and back tabs .
12 .  Replace deltoid injection pad on upper arm .

To remove arms(s), reverse procedure .

IV Arm:

Replacing Skin and Veins:

Laerdal Recommends

When excessive leaking occurs at the puncture sites, either a new vein 

system or skin should be installed to reduce loss of fluid. 

We recommend working at a basin when replacing skin and veins .

Replacing skin:

1 .  Cut off skin . This can be done with a sharp knife or scalpel . (Figure 6)
2 .  Discard skin . 

Replacing veins: (Keeping skin)

1.  Lubricate inside of skin with liquid detergent; let it flow down into the finger 

area .

2 .  Begin at top of arm and slowly pull skin down and remove from arm . Do not 

roll, as that will cause skin to bind . Thumb will detach with skin .

3 .  Remove tubing from track in mandrel . Glue may need to be scraped away .
4 .  Rinse and dry vein grooves well and swab with alcohol . Be sure to remove 

any excess glue .

English

Fig . 4

Photo 14

Fig . 5

Fig . 6

Содержание ALS Simulator

Страница 1: ...llyVitalSim ections for Use Instrucciones de Utilizaci n Instruzioni per l Uso Mode d Emploi Gebrauchsanweisung Handleiding Bruksanvisning Bruksanvisning K ytt ohje Nursing KellyVitalSim Nursing Kelly...

Страница 2: ......

Страница 3: ...also cause discoloration Care should be taken when palpating the pulses to not use excessive force as this will result in no pulse being felt Only two individual pulses can be felt at the same time If...

Страница 4: ...ertion Mid axillary site Left Cardiac Related Skills 3 4 lead ECG with over 1100 rhythm variations available for interpretation Pacing with variable threshold with or without capture Defibrillation ca...

Страница 5: ...livered with a set of pupils containing normal pupils mounted in the head A separate case contains 3 sets of plastic pupil inserts normal constricted and dilated pupils for use in simulating other con...

Страница 6: ...Note A blood pressure cuff can be used as a pressure infuser This will allow the blood bag to be placed on the floor or between the mattress and frame Photo 6 7 When the blood source bag is empty swit...

Страница 7: ...ctual voltages and current used on a live patient All precautions and safety measures must be used during the defibrillation and pacing phases of training Failure to follow safety measures could resul...

Страница 8: ...llar is showing wear discard and replace with a new collar IV Cannulation After IV training is complete 1 Close both clamps on IV tubing and detach bag tubing from arm tubing Fill a 100cc syringe with...

Страница 9: ...ads allow them to dry then store or reinsert in manikin 5 Articulating parts will benefit from a light application of talcum powder prior to training sessions To extend the cleanliness of your manikin...

Страница 10: ...ession 380455 Skin Chest 380460 Genitalia Blank Adult Male 380461 Kit Female Genitalia W Urinary Valves 380462 Kit Male Genitalia W Urinary Valves 380470 Belly plate Assembly 380471 Plug Belly plate A...

Страница 11: ...er Chestrise 205 01750 Thigh Tank Hardware Kit 205 10101 Manual Air Pump 383110 Chest Tube Module Replacements 200 03050 Pupil Set Replacement 200 03150 Airway Tongue Assembly 205 09001 Peripheral Kit...

Страница 12: ......

Страница 13: ...mientras manipule el maniqu podr a decolorarse en las zonas en las que lo haya tocado Tenga cuidado de no aplicar una presi n excesiva al tomar el pulso al maniqu en los distintos sitios porque ello...

Страница 14: ...troducci n del tubo tor cico Regi n media axilar izquierda Habilidades relacionadas con aspectos card acos ECG de 3 4 derivaciones con m s de 1100 variaciones card acas diferentes disponibles para su...

Страница 15: ...n de las pupilas El Simulador SVA est equipado con un juego de pupilas formado por pupilas normales colocadas en la cabeza En una caja independiente hay tres juegos de pupilas de pl stico insertables...

Страница 16: ...arterial a modo de infusor de presi n De este modo la bolsa de sangre se podr colocar en el suelo o entre el colch n y el gotero Fotograf a 6 7 Cuando la bolsa de sangre est vac a col quela donde est...

Страница 17: ...rilaci n y estimulaci n deben adoptarse todas las precauciones y medidas de seguridad oportunas En caso de no seguir estas medidas de seguridad pueden producirse lesiones o incluso la muerte de los us...

Страница 18: ...una nueva Canulaci n para tratamiento intravenoso Una vez finalizada la ense anza para el tratamiento intravenoso 1 Cierre las dos pinzas de los tubos para tratamiento intravenoso y separe de los tubo...

Страница 19: ...o antes de las sesiones de formaci n Si desea mantener siempre limpio el maniqu 1 Utilice una peque a toalla y aplique con cuidado una peque a cantidad de talco incluido a la cabeza y el t rax del man...

Страница 20: ...de clavijas ECG desfibrilador 205 03650 Manguito de presi n 380455 Piel del t rax 380460 Genitales de hombre adulto completo 380461 Juego de genitales femeninos con v lvulas urinarias 380462 Juego de...

Отзывы: