18

6.7 Replace the Battery

Note:

Use only batteries recommended by Laerdal.

1. Open the Battery Door.

2. To remove the battery, push and move it

slightly to the left and then release.

3. Withdraw the battery from the LSU.*

4. To insert a battery, push it fully in and then 

to the right to lock it.

5. Close the Battery Door.

6. After inserting the battery, place the LSU on 

charge unless a fully charged battery 

is inserted.

*When you discard the battery, dispose of safely in

accordance with local protocols for sealed lead acid

batteries.

6.8 Fastening brackets 

The fastening bracket is used to hold the LSU in

the optional wall bracket. Inspect the fastening

brackets for wear and tear regularly.

Replace if worn. See section 8.

6.9 Release arm for Canister holder 

An extra release arm for the canister holder can

be attached. See section 8.

7 Device test

The Device test is user initiated test program for

all models which can identify whether the LSU

operates satisfactorily or if the LSU needs 

service.

The program runs 4 different tests:

1 It tests for occlusions in the Suction System 

(including canister and tubing)

2 It tests the vacuum build-up efficacy of the 

Pump System (How much vacuum is built up 

within 3 seconds)

3 It tests the maximum achievable vacuum level 

of the LSU (reaches this level within 

10 seconds)

4 It tests if there are air leakages in the Pump 

System (including canister and tubing)

7.1 Setup

Before the test can be performed 

make sure that:

-  The Patient Suction Tubing is unwound and 

not blocked or bent 

- The Suction Catheter Adapter is removed 

from its holder (if applicable)

- The canister lid,T-bar, angled connector 

and tubing are securely fastened.

7.2 Run the Test

1 Press and hold the TEST-Button while setting 

the Operating Knob to 500+ mmHg.

Note: Do not release the TEST-Button until min.

2 seconds after the Operating Knob has been set

to 500+ mmHg. The test will start immediately

Fastening 

brackets

Содержание 78 00 00

Страница 1: ...Directions for Use Laerdal Suction Unit www laerdal com 78 00 10 78 00 00...

Страница 2: ...le to place the LSU on continuous charge when not in use make sure the battery is charged for a minimum of 24 hours at least once a month Do not store the battery when it is discharged always fully ch...

Страница 3: ...lace the Disposable Canister 13 5 3 Assembly of the Disposable Canister System 14 5 4 After Use 15 6 Maintenance 15 6 1 General 15 6 2 Battery 15 6 3 Cleaning the Cabinet 15 6 4 Decontamination of Reu...

Страница 4: ...m levels are usually selected for tracheal suctioning and the suctioning of children and infants 2 3 Contraindications None known 2 4 Features and Benefits Splashproof rugged and solid Easy access to...

Страница 5: ...nt Suction Tubing Suction Catheter Adapter Holder for Water Bottle Patient Suction Tubing Water Bottle Canister Lid with Float Ball Battery Door Release Arm Canister 3 1 2 LSU w Disposable Canister Sy...

Страница 6: ...it continuously while the LSU is switched ON It flashes rapidly approx twice per second during the Device Test It flashes slowly approx once per second while the Automatic Power save Function is activ...

Страница 7: ...ately after switching from external power to internal battery operation all 4 LED s will flash for 5 seconds before the remaining battery capacity is displayed CAPACITY for a new battery stored in roo...

Страница 8: ...d without the Aerosol Filter or the Float Ball It is recommended always to have extra Aerosol Filters with the LSU in case one has to be discarded The Aersol Filter is not intended for microbio logica...

Страница 9: ...ng external Power Sources AC mains when used with the AC Power Cord 100 240 VAC 50 60 400 Hz DC mains when used with the DC Power Cord 12 28 VDC A Wall Bracket to hold the LSU during operation and opt...

Страница 10: ...ting Knob to the required vacuum level The LSU will automatically be switched ON and start to operate The green Power ON Indicator is lit continuously while the LSU is switched ON 6 When suction is co...

Страница 11: ...yellow Gasket in the Lid make sure that the gasket is placed flat in the Lid and not twisted 2 If the Float Ball has been removed hold the lid upside down and snap the ball into place as shown Note Ne...

Страница 12: ...olor code Blue on Blue Yellow onYellow 4 3 1 Reusable Canister System Quick reference The assembly of the tubing is illustrated on the right side of the LSU 1 Connect the blue Aerosol filter to the bl...

Страница 13: ...erated in the upright position to prevent overflow of suctioned material 7 Set the Operating Knob to the required vacuum level and the LSU will automatically be switched ON and start to operate The gr...

Страница 14: ...to the blue Vacuum Inlet on the LSU Color Code Blue on Blue 4 Insert the Canister in the in the Canister Holder and slide it into the LSU 5 Connect the yellow Connector on the Vacuum tubing to the Ye...

Страница 15: ...ed both on a monthly basis and after each use 6 2 Battery To maintain satisfactory operation of the battery it is recommended to place the LSU on continuous charge when not in use Replace the battery...

Страница 16: ...eparate parts as indicated in the picture above The Float Ball on the reusable canister can be snapped out of the lid 6 4 3 Rinsing Rinse all parts under cold running water Immerse in warm water 30 40...

Страница 17: ...24 hours at least once a month Do not store the battery when it is discharged Always fully charge the battery before storage The recommended ambient temperature for charging is from 15 C to 25 C The...

Страница 18: ...is user initiated test program for all models which can identify whether the LSU operates satisfactorily or if the LSU needs service The program runs 4 different tests 1 It tests for occlusions in th...

Страница 19: ...est results turn the Operating Knob to 0 to exit the Device Test Note If you need to interrupt the test and revert to nor mal operation turn the Operating Knob to another position and then select the...

Страница 20: ...LED 2 lit up LED 3 lit up LED 4 lit up Test result indications Test Passed 100 mmHg Test Passed 300 mmHg Test Passed 500 mmHg Test Passed 450 mmHg Action if test failed Check possible blockages e g tw...

Страница 21: ...ter adapter pkg 10 78 40 00 LSU Reusable Canister 78 04 30 LSU Reusable Canister Holder 78 10 06 Angled Connectors pkg 10 78 10 02 Float Ball pkg 10 78 40 07 Vacuum Seal 78 40 08 Gasket 79 35 00 Water...

Страница 22: ...n the Abbott canister is leaking air into the tubes The Abbott liner is leaking Battery is not charged Tube s is kinked or twisted The external Power indicator is not lit ACTION Check power cord conne...

Страница 23: ...e Overall tolerance 5 11 3 Physical Characteristics Size 315 mm 12 4 in x 330 mm 13 in x 160 mm 6 3 in h x w x d Weight 4kg 8 9 lbs including battery Canister Capacity 1000 ml Canister graduation accu...

Страница 24: ...front Styrene Ethylene Betyl Styrene SEBS Cabinet back PC ABS Cabinet base w Protector PC ABS SEBS Battery door SEBS Connector retainer for battery Poly Oxy Methylene POM Operating Knob POM Rotor for...

Страница 25: ...rovided in this section Portable and mobile RF communication equipment can affect MEDICAL ELECTRICAL EQUIPMENT Warning MEDICAL ELECTRICAL EQUIPMENT should not be used adjacent to or stacked with other...

Страница 26: ...30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Mains power quality...

Страница 27: ...cy rangeb Interference may occur in the vicinity of equipment marked with the following symbol Guidance and manufacturer s declaration electromagnetic immunity NOTE 1 At 80 MHz and 800 MHz the higher...

Страница 28: ...2005 Laerdal Medical AS All rights reserved 8179 Rev A Printed in Norway 0434...

Отзывы: