12
8. Insert the Canister in the canister holder and
slide it into the LSU.
9. Connect the yellow Angled Connector to
the yellow Vacuum Seal
(Color code:Yellow on Yellow)
10. Connect the Angled Connector on the
Patient tubing to the PATIENT Inlet on the
canister Lid.
11. Perform the Device Test according to
section 7
12. Wind the Patient Tubing on the Patient
Tubing Holder
13. Place the LSU on charge (see section 6.5)
Note: Connect tubing according to color code:
Blue on Blue & Yellow on Yellow.
4.3.1 Reusable
Canister System Quick reference
The assembly of the tubing is illustrated on the
right side of the LSU.
1. Connect the blue Aerosol filter to the blue
Vacuum Inlet.
2. Connect the yellow angled connector to the
yellow Vacuum Seal.
3. Connect the Patient Tubing to the Patient inlet.
4.4 After Use
- When suction is complete, set Operating
Knob to ”0” and inspect all parts for damage
or excessive wear. Replace parts if necessary.
- Clean the LSU according to section 6.3 and its
reusable parts according to section 6.4.
- Perform the Device Test according
to section 7.
- Place the LSU on charge (see section 6.5).