6.4 Decontamination
of Laerdal Reusable Canister
6.4.1 Parts to be Decontaminated
After Each Use:
1. Canister Lid
2. Float Ball
3. Gasket.
4. Vacuum Seal
5. Angled Connectors
6. Canister
7. Patient Suction Tubing
8. Suction Catheter Adapter.
6.4.2 Preparation
- Remove and empty the Canister according
to section 4.2
- Disassemble the Canister into separate parts
as indicated in the picture above.
- The Float Ball on the reusable canister can
be snapped out of the lid.
6.4.3 Rinsing
- Rinse all parts under cold running water.
- Immerse in warm water (30-40°C).
6.4.4 Cleaning
- Immerse all parts in hot water (60-70°C)
containing a mild detergent (see section 6.3).
- Thoroughly clean all surfaces, use a brush
where possible.
- Rinse in warm water and allow to dry.
- Inspect all parts to be visibly clean and dry.
Note:
Thorough rinsing and cleaning are very important
steps prior to disinfection.
1.
4.
3.
8.
5.
7.
6.
2.