16
Prima ENT/DNT
ELECTRICAL CONNECTIONS
8
1
2
3
A
ON/OFF Switch
A/C Inlet
Connect the power cable to the AC inlet socket (1)
located at the back of the swivel arm (see Fig. 7).
Switch on the power using the on/off switch (2).
Note
: The power supply is designed with universal input
100V-240V AC, 50/60Hz. To plug in the power supply,
follow the instructions on the electrical label located at
the bottom of the arm (see 3 in Fig. 7).
1.
Replacing the Illumination Source
A. Detach the swivel arm covers.
B. Detach the fiber optic cable and replace the
illumination assembly (see A in Fig. 8) with a
new assembly.
C. Resecure the arm covers.
2.
Replacing the Fuse
A. Locate the fuse at the AC inlet (see Fig. 9).
B. Use a flat head screw driver to open the fuse
compartment. Two fuses are provided: one is
a live fuse and the second is a spare fuse.
C. Replace the blown fuse with the spare fuse
and close the fuse compartment.
For fuse replacement, refer to the label shown in
Fig. 9.
3.
Data Plate Details
UL Mark
Safety Certification Accompanying Document
Must Be Consulted
IEC60601-1:2005
Labeling for USA and Europe
IEC60601-1:2005
Labeling for Canada
Label Control
Number
Fuse Rating
Company
Trademark
Manufacturer
Separate Disposal
Waste Electric Device
Device Name
Model Number
Serial Number
Input Voltage
Frequency
Max Load
Fig 7
Fig 8
Fig 9