GUIDANCE TABLES (continued)
Prima is intended for use in the electromagnetic environment specified below. The customer or
user of the Prima should ensure that it is used in such an environment.
Guidance and Manufacturer’s Deceleration
Electromagnetic Emissions
All Equipment and Systems
Guidance and Manufacturer’s Deceleration - Electromagnetic Emissions
Immunity Test
Compliance Level
ESD
IEC 61000-4-2
± 6kv contact
± 8kv Air
± 2kv Mains
± 1kv I/Os
± 1kv Differential
± 2kv Common
± 1kv Differential
± 2kv Common
± 2kv Mains
± 1kv I/Os
± 6kv contact
± 8kv Air
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
Dropouts,
IEC 61000-4-11
Power Frequency
50/60Hz Magnetic
Field IEC 61000-4-8
3A/m
3A/m
IEC 60601 Test
Level
E
lectromagnetic Environment
- Guidance
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
Main power quality should be that of a
typical commercial or hospital Environ
-
ment. If the user of the Prima requires
continued operationduring power mains
interruption, it is recommended that the
Prima be powered from an uninterrupt-
able power supply or battery.
Floors should be wood, concrete or ceram-
ic tile. If floors are synthetic, the R/H
should be at least 30%.
Main power quality should be that of a
typical commercial or hospital environ
-
ment.
Main power quality should be that of a
typical commercial or hospital environ
-
ment.
Power frequency magnetic field should be
that of a typical commercial or hospital
Environment.
Prima
6137000-795
Prima
Issue 1.9
Printed on Sept,2021
63