Occupational safety and health protection:
Observe work safety and health protection of persons responsible for processing
contaminated products.
Current regulations of hospital hygiene and prevention of infection must be observed in
the preparation, cleaning and disinfection of the products.
Instructions
•
Workplace:
Remove surface contamination with a paper towel.
•
Reprocessing:
Recommended: reprocess a product immediately after use.
•
Cleaning& Servicing:
Needed: water, detergent, spirits, microfiber cloth
o
Flush the surface with running water (<40
˚
C), using a little detergent if necessary.
o
Also use spirits to clean optical components.
o
Dry optical components using a microfiber cloth; dry the rest of the product using a
paper towel.
o
Service as and when required should be informed to LABOMED after – sales service
department.
•
Autoclaving:
The asepsis sets available from LABOMED contain rubber caps, sleeves and grips
recommend for the following program for autoclaving:
Temperature:
134°
C
Time:
10
minute
Instrument:
Standard,
Autoclave
There are also rubber autoclavable caps for the rotating knobs for aperture
①
,
magnification
②
, green filter
③
, fine focusing
④
, tilt locking
⑤
, arm locking
⑥
, Ez-
lift
⑦
, and swivel arm locking
⑧
.
Autoclavable caps:
(C1) Part no. 6122015-209
(C2) Part no. 6168000-219
(C3), (C4) Part no. 6168000-213
(C5), (C6), (C7), (C8) Part no. 6168000-216