background image

 SL 20i 

TROUBLESHOOTING TABLE

16

Chart of Common Errors

ISSUE

PROBABLE CAUSE

POSSIBLE  SOLUTION

Lamp won’t turn on.

Incorrect input power supplied to
the SL 20i Slit Lamp.

Check the outlet to ensure proper power is
being supplied. 

Defective Power Cord.

Replace the Power Cord.

Bulb may be blown out.

Replace Bulb.

Defective Power Supply.

Replace the Power Supply.

Slit Lamp won’t move. Base Lock Screw may be tightened.

Rough base 
Movement.

Bearings may be damaged.

Replace the base bearing Assembly including
Shaft and Sleeve.

Replace the base bearing Assembly including
Shaft and Sleeve.

Shaft may be damaged.

Fixation light does not 
light up.

Fixation Light Harness not plugged 
into the Power Supply Assembly.

Ensure the Fixation Light Harness is properly 
seated in the Power Supply Assembly.

Defective Power Supply.

Replace the Power Supply.

Light too dim.

Incorrect wattage for bulb being used.

Bulb not installed properly.

Check bulb and ensure notch lines up with 
bulb housing.

Double slit visible in 
microscope.

Microscope not focused on focusing 
rod before use.

Install focusing rod and check to ensure 
microscope is focused on it.

Bulb not installed properly.

Check bulb and ensure notch lines up with
bulb housing. 

8126700-795 

 SL 20i 

Issue 1.0

Printed on June,2019

Loosen the Base Lock Screw.

Replace with the proper Bulb.

The following is a checklist of items that need to be assessed in order to determine if the SL 20i Slit Lamp
requires servicing.

       

   Check the outside of the Slit Lamp for any damage or missing components.

       

   Inspect the power cord for damage.

       

   Test the lamp by turning the lamp on and turning the light all the way to it’s brightest setting, and 

            all the way down to its lowest setting.
       

   Check to ensure all switches are functioning properly.

       

   Check the filters by cycling through all the options.

       

   Check the Slit Wheel by cycling through all the options.

       

   Check the base movement.

The following chart outlines some common issues with the SL 20i Slit Lamp and some steps you can
take to correct the issue. If problems persist, please contact the LABOMED as listed in the 

Introduction

section of this manual.

Содержание 8126700-795

Страница 1: ...Manual Part No 8126700 795 Issue 1 0 Printed on June 2019 To ensure proper use of this instrument as well as to avoid injury while operating Instrument understanding this manual completely before use...

Страница 2: ...Reserves the right to make change in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 13485 Certified LABOMED products are d...

Страница 3: ...ION OF INPUT POWER 11 INSTRUCTIONS FOR USE 12 OPERATION 12 SLIT LENGTH 13 FILTER DIAL 13 SLIT ROTATION 13 CLEANING MAINTENANCE 14 EXTERNAL CLEANING 14 FOREHEAD CHIN REST PREPARATION 14 CLEANING THE GL...

Страница 4: ...O ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING TO AVOID RISK OF ELECTRIC SHOCK THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAIN WITH PROTECTIVE EARTH OTHERWISE FAILURE OF INSTRUMENT AND...

Страница 5: ...AY OCCUR SEE THE CARE AND MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTIONS CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO ELECTROMAGNETIC CHARGE EMC AND NEEDS TO...

Страница 6: ...Power Protective Earth Connection ON OFF Manufacture Catalog Number Serial Number Waste of Electrical and Electronic Equipment Compliance to medical Device Directive 93 42 EEC Accompanying Document m...

Страница 7: ...ticians and eye care technicians should operate this instrument Please retain this manual for future reference and to share with other users Additional copies can be obtained from your authorized LABO...

Страница 8: ...k Knob 18 Instrument base Lock Knob 19 Guide Rail Cover 20 Geared Rollers 21 Guide Rails 22 Fore head support 23 Power Supply Assembly 24 Table Top 25 Patient Handles 26 Observation Head 27 Chin Suppo...

Страница 9: ...g Item Packed as Shown in Figure 2 a Chin Rest Assembly b Microscope Head with Eyepiece Mounted c Power Supply Box d Base Assembly e Patient Resting Handle f Illumination Arm Assembly g Accessories Sl...

Страница 10: ...ck the base at any position to prevent any undesired movement by using Base Locking Knob Shown as A in Fig 5 8 Assemble Illumination moving arm assembly onto the base using Allen Screw M6 as shown in...

Страница 11: ...MINER OR A DANGER TO THE PATIENT WARNING POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION DEVICE PLUG 1 After the unit is in its secure location apply the correct inpu...

Страница 12: ...bs adjust the projected slit so that the thinnest slit is shown on the Focusing Rod Refer to fig 12 15 NOTE The thinnest line will allow for greater accuracy 6 Remove the Focusing Rod 7 To position a...

Страница 13: ...dicated in the scale of the slit lamp arm Refer to figure 17 15 Magnification is altered by using lever on the Magnichanger Refer to figure 16 SLIT LENGTH The Slit length is adjusted by rotating the S...

Страница 14: ...dampened with a mild soap and water solution CHANGING THE HALOGEN BULB WARNING NEVER REMOVE A BULB THAT HAS RECENTLY BEEN IN USE AS IT WILL BE VERY HOT WAIT UNTIL IT HAS COOLED AND USE GLOVES OR A TH...

Страница 15: ...y FUSE REPLACEMENT Replace the fuses in the Power Input Module with the fuses indicated in the Specifications section of this manual 1 Remove input power to the instrument 2 Press down on the top tab...

Страница 16: ...bulb housing Double slit visible in microscope Microscope not focused on focusing rod before use Install focusing rod and check to ensure microscope is focused on it Bulb not installed properly Check...

Страница 17: ...80 Atmospheric Pressure 80 kPa 23 6 in Hg to 106 kPa 31 3 in Hg Transportation Storage 10 Temperature 20 C 4 F to 70 C 158 F 106 kPa Relative Humidity 10 to 80 non condensing Atmospheric Pressure 50k...

Страница 18: ...nnecting the equipment to a protective grounding system so that the accessi ble metal parts do not become conductive in case of failure in the basic insulation Grade of protection against a hazardous...

Страница 19: ...T H S ADAPTER KIT SBS WITH VIDEO OUTPUT CCD ADAPTER 8122500 8122600 800 6134131 6134110 6134120 6134172 I PHONE 6 6134172 1 6 I PHONE 6 PLUS 6134172 1 6P 8122600 851 8122600 852 CCD CAMERA I PHONE 7 6...

Страница 20: ...Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class A Harmonics IEC61000 3 2 Class A Flicker IEC61000 3 3 Complies 8126700 795 SL 20i Issue 1 0 Printed onJune 201...

Страница 21: ...5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds Power Frequency 50 60Hz Magnetic Field IEC61000...

Страница 22: ...ccording to the transmitter manufacturer and dis therecom mended separation distance in meters m Field strengths from fixed Rf transmitters as determined by an electromagnetic site survey should be le...

Страница 23: ...or use in the electromagnetic environment in which radiated RF disturbance are controlled The customer or user of the SL 20i can help prevent Electromagnetic interference by maintaining a minimum dist...

Страница 24: ...t Labomed shall not be liable for any special Incidental or consequent damages for any negligence breach of warranty strict liability or any other damages resulting from or relating to design manufact...

Страница 25: ...Revision History Rev No Date of Release Change App By DCR...

Страница 26: ...4538 U S A Phone 510 445 1257 Fax 510 991 9862 E mail sales laboamerica com www laboamerica com Labomed Europe Essebaan 50 NL 2908 LK Capelle a d IJssel The Netherlands Tel 31 0 10 4584222 Fax 31 0 10...

Отзывы: