Shipping damage is the responsibility of the transportation carrier.
Any claims for damage must be filed within 48 hours.

The accessories supplied with the centrifuge should be kept with the
instruction manual near the centrifuge’s place of installation.

2.2  Required space

The centrifuge should be installed on a rigid, even surface such as a
stable laboratory bench, cabinet, etc.  To guarantee sufficient ventila-
tion, ensure that the centrifuge has at least 15cm (6 inches) of free
space on all sides, including the rear.

The centrifuge should not be located in areas subject to excessive
heat such as in direct sunlight or near radiators or the exhaust of a
compressor, as a buildup of heat may occur within the chamber.

2.3  Installation

Before operating the centrifuge, check that the power source corre-
sponds to that on the manufacturer’s rating label, then connect the
power cord to the centrifuge and the power source.

3.  Installation or rotors and rotor maintenance

3.1  Rotors and accessories

The following accessories are available for the Spectrafuge:

Angle rotor for 6 x 15ml tubes

Order  no.

Included with unit (C0060-RTR)

Tube measurement

15ml (17x125mm)-10ml (16x100mm)

Max. speed

6,500rpm

Centrifuging radius

8.5cm

RCF (g-value)

4,000 x g

Adapter for 5ml(12x75mm) and 7ml (13x100mm) tubes

Order no.

C0200-17A

Tube measurement

12x75mm, 13x100mm, or commom
Sarstetd style tubes

3

4.5  Selection of operating time and momentary operation

Time can be selected in half minute intervals from 0.5 to 10
minutes, and in one minute intervals from 11 to 30 mintes.  Time
can also be set to continouos/hold by turning the timer knob past
the thirty minute position.  This will display the continouos setting
“On”.

When the preselected time expires, the centrifuge will stop
automatically.  To stop the centrifuge prior to the expiration of set
time, press the “start/stop” knob.

The centrifuge may be operated manually by pressing and holding
the “start/stop” knob.  The centrifuge will continue to run as long
as the knob is depressed.

4.6  Starting the centrifuge

Once the time and speed have been set the centrifuge can be
started by pressing the “start/stop” knob.  The centrifuge will the
run for the specified amount of time.

5.  Service and Maintenance

5.1  Centrifuge service

The Spectrafuge 6C requires no routine maintenance other than
the occasional routine cleaning.  All repairs should be performed
by authorized, qualified personnel only.  Repairs performed by
unauthorized personnel may void the warranty.

5.2  Cleaning the centrifuge

Always keep the centrifuge housing, rotor chamber, rotor and
rotor accessories clean.  All parts should be wiped down periodi-
cally with a soft cloth.  For more thorough cleaning, use a neutral
cleaning agent (pH between 6 and 8) applied with a soft cloth.
Excessive amounts of liquid should be avoided.  Liquid should not
come into contact with the motor.  After cleaning, ensure that all
parts are dried thoroughly by hand or in a warm air cabinet
(maximum temperature 50ºC)

8

Содержание Spectrafuge 6C

Страница 1: ...rafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Compact Centrifuge Labnet International PO Box 841 Woodbridge NJ 07095 Phone 732 417 0700 Fax 732 417 1750 Email labnet labnetlink...

Страница 2: ...rotor area unless the rotor is completely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flamma...

Страница 3: ...ed selection Selection of operating time and momentary operation Starting the Centrifuge 5 Service and Maintenance 8 Service Cleaning Disinfection Replacing fuses 6 Troubleshooting 9 7 Where to Call 1...

Страница 4: ...properly balanced 1 Possible reason Lid lock is jammed Solution Call for service Possible reason Centrifuge is not receiving power Solution See Centrifuge will not start Centrifuge cannot be started a...

Страница 5: ...aning the rotor The rotor should be cleaned after each use When spinning samples containing phenol or phenol chloroform the rotor should be cleaned immediately after use 5 4 Disinfection Should a spil...

Страница 6: ...of operating time and momentary operation Time can be selected in half minute intervals from 0 5 to 10 minutes and in one minute intervals from 11 to 30 mintes Time can also be set to continouos hold...

Страница 7: ...on the large LED display right Figure 4 Spectrafuge controlpanel layout 7 Tube and tube adapter reference chart 3 2 Rotor maintenance The rotor should be cleaned thoroughly after each use Thor ough cl...

Страница 8: ...sult in damage to the rotor and centrifuge The revised maximum speed can be calculated with the following formula Reduced speed nred x max speed nmax Example Where the density of the liquid is 1 7 the...

Отзывы: