1.3  Technical data

Dimensions
Width

8.25 inches

Depth

9.5 inches

Height

7.0 inches

Maximum speed

6,500rpm

Maximum RCF

4,000 x g

Maximum volume

6 x 15ml

Admiss. density

1.2kg/dm

3

Electrical/fuse rating

100V~,50-60Hz,  1.0A/1.6 AT

120V~, 50-60Hz, 1.0A/2.5AT
230V~, 50-60Hz, 0.6A/1.25AT

1.4  Accessories supplied with centrifuge

Each unit is supplied with 1 instruction manual, 1 warranty card
and 1 power cord.  Some models are supplied with a rotor screw
wrench.

1.5  Warranty

This centrifuge has been subject to thorough testing and quality
control.  In the unlikely event of a manufacturing fault, our one
year warranty (from the date of delivery) covers the centrifuge
and the rotor.  This warranty becomes invalid in the case of
incorrect operation, use of nonstandard spare parts or accesso-
ries and unauthorized modification of the rotor or centrifuge.

Labnet reserves the right to make technical modifications.
Statements contained herein are not to be considered binding.

2.  Installation

2.1  Unpacking the centrifuge

Before unpacking the centrifuge, inspect the outside of the carton
for any shipping damage.

The centrifuge is delivered in a carton with protective cushions.
Remove the centrifuge from the carton.  Retain the carton and
cushions until it has been established that the centrifuge is
working properly.

Inspect the centrifuge for any visible signs of shipping damage.

2

5.3  Cleaning the rotor

The rotor should be cleaned after each use.  When spinning
samples containing phenol or phenol chloroform, the rotor should
be cleaned immediately after use.

5.4  Disinfection

Should a spill of infectious materials occur within the rotor or
chamber, the unit should be disinfected.  This should be per-
formed by qualified personnel with proper protective equipment.

5.5  Replacing fuses

Check the fuse when it is recommended in the Troubleshooting
Guide located in this manual.  The fuse holder is located in the
power inlet on the rear of the unit.  Disconnect the power cord
from the power inlet.  Open the fuse holder drawer by inserting a
small screwdriver under the tab and prying it open.  Remove the
innermost (operative) fuse from its retaining tabs and replace the
fuse if necessary.  A spare fuse is located in the outermost
chamber of the fuse drawer.  Replace only with a fuse of exactly
the same value as the original.  (Fuse type may be found in the
Technical data section of this manual.)

6.  Troubleshooting Guide

Please refer to this guide before calling for service.

Centrifuge will not start

Possible reason:

No power supply

Solution:

Check that power is being supplied to
the outlet
Check that the power cord is plugged
into both the wall outlet and the back of
the centrifuge
Check that power cord is not damaged

Possible reason:

Blown fuse

Solution:

Check fuse and replace if necessary

Lid lock will not release

Possible reason:

Defective lid lock

Solution:

Open manually and have unit serviced

Possible reason:

No power from PC board

Solution:

Call for service

9

Содержание Spectrafuge 6C

Страница 1: ...rafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Compact Centrifuge Labnet International PO Box 841 Woodbridge NJ 07095 Phone 732 417 0700 Fax 732 417 1750 Email labnet labnetlink...

Страница 2: ...rotor area unless the rotor is completely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flamma...

Страница 3: ...ed selection Selection of operating time and momentary operation Starting the Centrifuge 5 Service and Maintenance 8 Service Cleaning Disinfection Replacing fuses 6 Troubleshooting 9 7 Where to Call 1...

Страница 4: ...properly balanced 1 Possible reason Lid lock is jammed Solution Call for service Possible reason Centrifuge is not receiving power Solution See Centrifuge will not start Centrifuge cannot be started a...

Страница 5: ...aning the rotor The rotor should be cleaned after each use When spinning samples containing phenol or phenol chloroform the rotor should be cleaned immediately after use 5 4 Disinfection Should a spil...

Страница 6: ...of operating time and momentary operation Time can be selected in half minute intervals from 0 5 to 10 minutes and in one minute intervals from 11 to 30 mintes Time can also be set to continouos hold...

Страница 7: ...on the large LED display right Figure 4 Spectrafuge controlpanel layout 7 Tube and tube adapter reference chart 3 2 Rotor maintenance The rotor should be cleaned thoroughly after each use Thor ough cl...

Страница 8: ...sult in damage to the rotor and centrifuge The revised maximum speed can be calculated with the following formula Reduced speed nred x max speed nmax Example Where the density of the liquid is 1 7 the...

Отзывы: