1.  General Information

This manual provides important safety information for the Spectrafuge 6C
laboratory centrifuge.  It should be kept near the centrifuge for quick and
easy reference.

1.1  Description

The Spectrafuge 6C is a small benchtop centrifuge designed for
separation of various research samples.  The Spectrafuge 6C is
supplied with an 6 x 15ml rotor.  Adapters are available for tubes
smaller than 10ml.  The Spectrafuge reaches speeds of up to
6,500rpm/4,000 x g.

1.2 Safety precautions

Note:  All users of the centrifuge must read the Safety
Precautions section of this manual before attempting to
operate the unit!

If this equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment
may be impaired.

Do not operate the centrifuge if any of the following conditions
exist:

-The centrifuge has not been installed properly
-The centrifuge is partially dismantled
-Service has been attempted by unauthorized or unqualified

personnel

-The rotor has not been installed securely on the motor shaft
-Rotors and accessories not belonging to the standard range are

being used without permission being obtained from the manu-
facturer to use such rotors and/or accessories in the centrifuge

 Exception:  Centrifuge tubes, normally available in the laboratory.
-The centrifuge is located in an explosive atmosphere
-Materials to be centrifuged are combustible and/or explosive
-Materials to be centrifuged are chemically reactive
-The rotor load is not properly balanced

!

1

Possible reason:

Lid lock is jammed

Solution:

Call for service

Possible reason:

Centrifuge is not receiving power

Solution:

See “Centrifuge will not start”

Centrifuge cannot be started, although power is on

Possible reason:

Lid not closed correctly

Solution

             Close lid correctly

Possible reason:

No speed or time has been selected

Solution:

Set speed and/or time

Centrifuge displays error “03”

Possible reason:

Lid opened prior to  signal “00”.

Solution

Close lid and open lid

7.  Where to call

Should you have any questions about the Spectrafuge or its accesso-
ries, please call Labnet’s Customer Service Department at 732 417-
0700.  Customer Service is staffed from 8:30am to 5:30pm, EST,
Monday through Friday.  Our 24 hour fax number is 732 417-1750.
Inquiries may also be sent via our electronic mailbox at
[email protected].

Should your Spectrafuge require service, please call Labnet’s Techni-
cal Services Department at 732 417-0700.  Our Service Department
is staffed from 8:30am to 5:00pm, EST, Monday through Friday.  Our
24 hour fax number is 732 417-1750.  Electronic mail may be sent to
[email protected].

Please have the unit’s serial number (located on the back panel of the
instrument) available when calling.  Should an item require return to
Labnet for service, a repair return authorization (RRA) number must
first be received from Labnet.  Items sent without an RRA number
will not be accepted.

8.  Determination of g-values

The centrifuging radius of the 15ml rotor is 8.5cm for round bottom
tubes and 8.2 for conical bottom tubes.  See Section 3.1 for the
correct radius when using adapters and smaller tubes.  The chart on
the next page can be used to determine g-values.

10

Содержание Spectrafuge 6C

Страница 1: ...rafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Compact Centrifuge Labnet International PO Box 841 Woodbridge NJ 07095 Phone 732 417 0700 Fax 732 417 1750 Email labnet labnetlink...

Страница 2: ...rotor area unless the rotor is completely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flamma...

Страница 3: ...ed selection Selection of operating time and momentary operation Starting the Centrifuge 5 Service and Maintenance 8 Service Cleaning Disinfection Replacing fuses 6 Troubleshooting 9 7 Where to Call 1...

Страница 4: ...properly balanced 1 Possible reason Lid lock is jammed Solution Call for service Possible reason Centrifuge is not receiving power Solution See Centrifuge will not start Centrifuge cannot be started a...

Страница 5: ...aning the rotor The rotor should be cleaned after each use When spinning samples containing phenol or phenol chloroform the rotor should be cleaned immediately after use 5 4 Disinfection Should a spil...

Страница 6: ...of operating time and momentary operation Time can be selected in half minute intervals from 0 5 to 10 minutes and in one minute intervals from 11 to 30 mintes Time can also be set to continouos hold...

Страница 7: ...on the large LED display right Figure 4 Spectrafuge controlpanel layout 7 Tube and tube adapter reference chart 3 2 Rotor maintenance The rotor should be cleaned thoroughly after each use Thor ough cl...

Страница 8: ...sult in damage to the rotor and centrifuge The revised maximum speed can be calculated with the following formula Reduced speed nred x max speed nmax Example Where the density of the liquid is 1 7 the...

Отзывы: